{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63407",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Road",
      "address_2": "N/A",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AL , CA , CO , DC , DE , FL , GA , IL , IN , KS , KY , LA , MA , MI , MO , MS , MT , NC , ND , NE , NH , NY , OH , PA , TN , TX , VA , and WI.  Internationally to  Australia, United Arab Emirates (UAE), Singapore,  Malaysia, BELGIUM, Japan, and CANADA.",
      "recall_number": "Z-0461-2013",
      "product_description": "24 Fr 8mm Soft Flow Ang Wire W/L    The Sarns Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.",
      "product_quantity": "5760 total products",
      "reason_for_recall": "The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl",
      "recall_initiation_date": "20121121",
      "center_classification_date": "20121203",
      "termination_date": "20130423",
      "report_date": "20121212",
      "code_info": "catalog number: 5762 and lot number: 0570466 and 0589212"
    }
  ]
}