      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97665",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.",
      "recall_number": "Z-0452-2026",
      "product_description": "Brand Name: Lithotriptor, Ultrasonic  Product Name: ShockPulse Lithotripsy Transducer  Model/Catalog Number: SPL-T  Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.  Component: ShockPulse Lithotripsy System",
      "product_quantity": "1528 units",
      "reason_for_recall": "The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.",
      "recall_initiation_date": "20251008",
      "center_classification_date": "20251107",
      "report_date": "20251119",
      "code_info": "Model/Catalog Number: SPL-T; UDI-PI: 00855279005023; All serial numbers;",
      "more_code_info": ""
    }
  ]
}