  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97882",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;",
      "recall_number": "Z-0448-2026",
      "product_description": "Convenience Kits:  DYNJ901048N\tORTHO SPINE ST CLAIR  DYNJ909123F\tUROLOGY MINOR  DYNJ909132F\tSET UP  DYNJ909308D\tHIP PINNING",
      "product_quantity": "96,406 total units",
      "reason_for_recall": "Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120\" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.",
      "recall_initiation_date": "20251010",
      "center_classification_date": "20251113",
      "report_date": "20251119",
      "code_info": "HIP PINNING DYNJ909308D UDI-DI 40198459055288 (case), 10198459055287 (ea) Lots 25FBB311 25GBC282 25GBW486;  ORTHO SPINE ST CLAIR DYNJ901048N UDI-DI 40198459055554 (case), 10198459055553 (ea) Lots 25GBQ070;  SET UP DYNJ909132F UDI-DI 40198459130572 (case), 10198459130571 (ea) Lots 25GBI845 25IBF708;  UROLOGY MINOR DYNJ909123F UDI-DI 40198459130602 (case), 10198459130601 (ea) Lots 25GBT826 25IBF707",
      "more_code_info": ""
    }
  ]
}