  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Horsham",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89092",
      "recalling_firm": "Flower Orthopedics Corporation",
      "address_1": "100 Witmer Rd Ste 280",
      "address_2": "N/A",
      "postal_code": "19044-2647",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, CA, CO, GA, TN, FL, IL, MA, MD, NC, TX.",
      "recall_number": "Z-0448-2022",
      "product_description": "Flex-Thread Inserter Shaft intended for  Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue Ratchet Handle in order to facilitate the insertion of the Flex-Thread Nail System-  Catalog Number: 8435-1",
      "product_quantity": "48 units",
      "reason_for_recall": "Mating incompatibility between the Inserter Shaft and the Inserter Draw Rod causing the inserter draw rod to bind and not rotate freely as intended surgeon may be unable to complete surgery unless a second set is available",
      "recall_initiation_date": "20211102",
      "center_classification_date": "20220103",
      "termination_date": "20231012",
      "report_date": "20220112",
      "code_info": "Lot Number: 11090 UDI: 00840118117548"
    }
  ]
}