  "results": [
    {
      "status": "Ongoing",
      "city": "Elyria",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89062",
      "recalling_firm": "Invacare Corporation",
      "address_1": "1200 Taylor St",
      "address_2": "N/A",
      "postal_code": "44035-6248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.",
      "recall_number": "Z-0445-2022",
      "product_description": "Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A)  Models: TDX SP2: TDXSP2, TDXSP2-CG, TDXSP2X-CG, TDXSP2-MCG, TDXSP2X-MCG, TDXSP2V, TDXSP2V-HD, TDXSP2HD, TDXSP2HD-CG, TDXSP2HD-MCG",
      "product_quantity": "1477 units US; 1226 units OUS",
      "reason_for_recall": "Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury",
      "recall_initiation_date": "20211112",
      "center_classification_date": "20220103",
      "report_date": "20220112",
      "code_info": "LiNX Gyro firmware version 6.1.2  UDI-DI Numbers:       TDXSP2  00841447100911     TDXSP2-CG  00841447100942     TDXSP2-MCG  00841447100966     TDXSP2HD  00841447108641     TDXSP2HD-CG  00841447111207     TDXSP2HD-MCG  00841447111214     TDXSP2V  00841447101512     TDXSP2V-HD  00841447102052     TDXSP2X-CG  00841447101000     TDXSP2X-MCG  00841447101017"
    }
  ]
}