      "country": "United States",
      "city": "Portage",
      "address_1": "3800 E Centre Ave",
      "reason_for_recall": "The beds with the high rail side rail option may not be complaint with IEC 60601-2-52/C1:2010.  There is a potential risk of entrapment.",
      "address_2": "",
      "product_quantity": "20,030 total",
      "code_info": "Serial Numbers: 026192  026510  026511  026516  026517  026518  034095  034096  034097  034098  034099  034517  035212  035213  035214  035215  035216  035217  035218  035219  035220  035221  035222  035223  035224  035225  035226  036935  036936  036937  036938  036939  036940  036941  037995  037996  037997  037998  037999  038000  038001  038002  038003  038004  038005  038006  038126  038127  038128  039790  039791  041998  042590  044486  045322  045910  045911  045912  045913  045914  045915  045916  047581  048338  048339  048340  048341  048342  048344  048345  048346  048347  048348  034424  034425  034518  034519  034520  034521  034522  034523  034524  034525  034526  034527  034528  034529  039359  039447  039448  039449  039450  039451  039452  041416  041417  041418  041419  041420  041421  041422  041423  041424  041425  041870  041871  041872  041873  041874  041875  041876  041877  041878  041879  041880  041881  041882  041883  041884  041885  042250  042251  042252  042253  042254  042255  042256  042257  042258  042259  044754  038909  039561  044713  044714  044715  044716  044717  044718  044719  044720  044721  044722  044723  044724  044725  044726  044727  044728  044729  044730  044731  044732  044733  044734  044735  044736  044737  044738  044739  044740  022903  022904  022905  034445  034446  035133  035134  035135  035136  035137  035138  035139  035140  035141  035142  035143  035144  035145  035146  035147  035148  035149  035150  038726  038727  038728  038729  038730  038731  038732  038733  038734  038735  038736  038737  038738  038739  038740  038741  038742  038743  038744  038745  038746  038747  038748  038749  038750  038751  038752  038753  027260  027261  027262  027263  027264  022224  022225  022226  022227  022228  022229  022231  022232  022233  022613  022614  022615  022616  022617  022618  022619  022620  022621  022622  022623  022624  022625  022626  022627  022628  022629  022630  022631  022632  022633  022634  022635  022636  022637  022638  022639  022640  022641  022642  022643  022644  022645  022646  022647  022648  022649  022650  022651  022652  022653  022654  022655  022656  022657  022658  022659  022660  022661  022662  022663  022664  022665  022666  022667  022668  022669  022670  022671  022672  022673  022674  022675  022676  022677  022678  022679  022680  022681  022682  022683  022684  022685  028406  028407  028408  028409  028410  028411  028412  028413  028414  028415  028416  028417  028418  028419  028420  028421  028422  028423  028424  028425  028426  028427  028428  028429  028430  028431  028432  028433  028434  028435  028436  028437  028438  028439  028440  028441  028442  028443  028444  028445  028446  028447  028448  028449  028450  028451  028452  028453  028454  028455  028456  028457  028458  028459  028460  028461  028462  028463  028464  028465  028466  028467  028468  028469  028470  028471  028472  028473  028474  028475  028476  028477  028478  028479  028480  028481  028482  028483  028484  028485  028486  022221  022222  022223  032470  032471  034616  034617  034618  034619  034620  041490  041658  041659  041660  041661  041686  041687  041688  041689  041690  041691  041692  041693  041694  041695  041696  041697  041698  041699  041700  041701  041702  041703  041704  041705  041706  041707  041708  041709  041710  041711  041712  041713  041714  041715  041716  041717  041718  041719  041720  041721  041722  041723  041724  041725  041726  041727  041728  041729  041730  041731  041732  041736  041737  041738  041739  041740  041741  041742  041743  041744  041745  041746  041747  041748  041749  041750  041751  041752  041753  041754  041755  041756  041757  041758  041759  041760  041761  041762  041763  041764  041765  041766  041767  041768  041769  041770  041771  041772  041773  041774  041775  041776  041777  041778  041779  041780  041781  041782  041783  041784  041785  041786  041787  041788  041789  041790  041791  041792  041793  041794  041795  041796  041797  041798  041799  041800  041801  041802  041803  041804  041805  041806  041807  041808  041809  041810  041811  041812  041813  041814  041815  041816  041817  041818  041819  041820  041821  041822  041823  041824  041825  041826  041827  041828  041829  041830  041831  041832  041833  041834  041835  041836  041837  041838  041839  041840  041841  041842  041843  041844  041845  041846  041847  041934  042591  042592  042593  044334  045325  045326  045327  045328  046794  046795  046796  046797  046798  047147  047582  047583  047713  047714  048498  048499  048500  048501  048502  048503  048504  022214  022215  022216  022217  022218  022219  022220  024303  026315  026316  026514  034421  041933  047897  047898  048041  048042  048043  048044  048045  048046  048047  048048  048049  048050  048051  048052  048053  048054  048055  048056  048057  048058  048059  048060  048061  048062  048063  048064  048065  048066  048067  048068  048069  048070  048071  048072  048073  048074  048075  048076  048077  048078  048079  048080  048081  048082  048083  048084  048085  048086  048087  048088  048089  048090  048091  048092  030970  030971  030972  030973  030974  030975  030976  030977  030978  030979  030980  030981  030982  030983  030984  040881  040882  040883  040884  040885  040886  040887  040888  040889  040890  040891  040892  040893  040894  040895  040896  040897  040898  040899  040900  046458  047145  024359  024360  024361  024362  024363  024364  024365  024366  024367  024368  024369  024370  031128  031129  031130  031131  031132  031133  031134  031135  031136  031137  031138  031139  031140  031141  031142  031143  031144  031145  031146  031147  031148  031149  031150  031151  031152  031153  031154  031155  031156  031157  031158  031159  031160  031161  030296  030297  030298  030299  030300  030301  030302  030303  030304  030305  030306  030307  030308  030309  030310  030311  030312  030313  030314  030315  030316  030317  030318  030319  030320  030321  030322  030323  030324  030325  030326  030327  030328  030329  030330  030331  030332  030333  030334  030335  030336  030337  030338  030339  030340  030341  030342  030343  030344  030345  030346  030347  030348  030349  030350  030351  030352  030353  026193  027085  027086  027087  027088  027089  027090  027091  027092  027093  027094  027095  027096  027097  027098  027099  027100  027101  027102  027103  027104  027105  027106  027107  027108  027109  027110  027111  027112  027113  027114  027115  027116  027117  027118  027119  027120  027121  027122  027123  027124  027125  027126  027127  027128  027129  027130  027131  027132  027133  027134  027135  027136  027137  027138  027139  027140  027141  027142  027143  027144  027145  027146  027147  027148  027149  027150  027151  027152  027153  027154  027155  027156  027157  027158  027159  027160  027161  027162  027163  027164  027165  027166  027167  027168  027169  027170  027171  027172  027173  027174  027175  027176  027177  027178  027179  027180  027181  027182  027183  027184  027185  027186  027187  027188  027189  027190  027191  027192  027193  027194  027195  027196  027197  027198  027199  027200  027201  027202  027203  027204  027205  027206  027207  027208  027209  027210  027211  027212  027213  027214  028042  022264  022265  022266  022267  022268  022269  022270  022271  022272  022273  022274  022275  022276  022277  022278  022279  022280  022281  022282  022283  022284  022285  022286  022287  022288  022289  022290  022291  022292  022293  022294  022295  022296  022297  022298  022299  022300  022301  022302  022303  022304  022305  022306  022307  022308  022309  022310  022311  029878                     026501  026502  026503  026504  026505  026506  026507  026508  026509  026512  038417                     042482                     036504                     036505                     036506                     036507                     037394                     039147                     039148                     037993                     037994                     032190                     032191                     032192                     032193                     045217                     045218                     039738                     039739                     040083                     040084                     040085                     040086                     040087                     040088                     040089                     040090                     040091                     040092                     041967                     041968                     034038                     034039                     042738                     042739                     046938                     046939                     046940                     046941                     046942                     046943                     046944                     046945                     046946                     046947                     046948                     046949                     046950                     046951                     046952                     046953                     046954                     046955                     046956                     046957                     046958                     023016  023017  023018  023019  023020  023021  023022  023023  023024  023025  023026  023027  023028  023029  023030  023031  021797                     021798                     021799                     021800                     021801                     021802                     021803                     021804                     021805                     021806                     021807                     021808                     021809                     021810                     021811                     021812                     021813                     021814                     021815                     021816                     021817                     021818                     021819                     021820                     021821                     021822                     021823                     021824                     021825                     021826                     021827                     021828                     021829                     021830                     021831                     021832                     021833                     021834                     023851  033885                     033886                     033887                     033888                     033889                     033890                     033891                     033892                     033893                     033894                     033895                     033896                     033897                     033898                     033899                     033900                     033901                     033902                     033903                     033904                     033906                     033907                     033908                     033909                     033910                     033911                     033912                     033913                     033914                     033915                     033916                     033917                     037953                     037954                     037955                     037956                     037957                     037958                     037959                     037960                     037961                     037962                     037963                     037964                     033356                     033357                     033358                     033359                     033360                     033361                     033362                     033363                     033364                     033365                     033366                     033367                     033368                     033369                     033370                     033371                     033372                     033373                     039225                     039226                     039227                     039228                     039229                     039230                     039231                     039232                     039233                     039234                     039235                     039236                     039237                     039238                     039239                     039240                     039241                     039242                     039243                     039244                     039245                     039246                     039247                     039248                     039249                     039250                     039251                     039252                     039253                     039254                     039255                     039256                     039257                     039258                     039259                     039260                     035229                     035230                     035231                     035232                     035233                     035234                     035235                     035236                     035237                     035238                     035239                     035240                     035241                     035242                     035243                     035244                     042700                     042701                     042702                     042703                     042704                     042705                     042706                     042707                     035227                     035228                     044969                     044970                     044934                     042771                     042772                     042773                     042774                     042775                     042776                     047173                     047174                     047175                     047176                     047177                     047178                     047179                     047180                     047181                     047182                     047183                     047184                     047185                     047186                     047187                     047188                     047189                     047190                     047191                     047192                     047193                     047194                     047195                     047670                     047671                     047672                     044187                     044188                     044189                     044190                     044191                     044192                     044114                     044115                     044116                     044117                     046295                     046296                     041868                     041869                     039299                     042594                     042595                     042600                     042602                     033107                     039562                     039563                     039564                     039565                     039566                     039567                     039640                     039641                     039642                     039643                     039644                     039645                     032738                     032739                     032740                     032741                     032742                     032743                     032744                     032745                     044335                     044336                     044337                     044338                     044339                     044340                     044341                     044342                     044343                     044344                     044345                     044346                     044347                     044348                     044349                     044350                     046888                     039206                     039207                     039208                     039209                     039210                     039211                     039212                     039213                     039914                     039915                     046297                     046298                     047355                     047356                     047357                     047358                     047359                     047360                     042477                     042478                     042479                     042480                     042481                     039911                     039912                     034100                     034101                     034102                     034103                     034104                     034105                     034106                     034107                     034108                     034109                     034110                     034111                     034112                     034113                     034114                     034115                     034116                     034117                     034118                     034119                     034120                     034121                     034122                     034123                     034124                     034125                     034126                     034127                     034128                     034129                     034130                     034131                     034132                     034133                     034134                     034135                     034136                     034137                     034138                     034139                     039069                     039070                     035324                     035325                     035326                     035327                     035328                     035329                     035330                     035332                     035333                     035334                     035335                     035336                     035337                     035338                     035339                     035340                     035341                     035342                     035343                     037783                     035331                     038669                     038670                     038671                     038672                     038673                     038674                     038675                     038676                     038677                     038678                     038679                     042365                     042366                     042367                     042368                     042369                     042370                     042371                     042372                     042373                     042374                     042375                     042376                     040864                     032131                     038910                     038911                     038912                     038913                     040559                     040560                     040561                     042768                     042769                     042770                     046713                     046714                     046715                     046716                     044560                     044561                     042790                     047326                     047327                     041142                     041143                     041144                     041145                     041146                     041147                     041148                     041149                     041150                     041151                     041152                     041153                     031716                     031717                     031718                     031719                     036948                     036949                     048096                     048097                     048098                     048099                     048100                     048101                     048102                     048103                     048104                     048105                     048106                     048107                     048108                     048109                     048110                     039062                     039063                     039064                     039065                     039066                     039067                     039068                     047103                     047104                     047105                     047106                     047107                     038354                     038355                     038356                     038357                     038358                     038359                     038360                     038361                     038362                     038363                     038364                     038365                     038366                     038367                     038368                     045631                     045632                     045633                     045634                     045635                     045636                     045637                     045638                     045639                     045640                     045641                     045642                     045643                     045644                     045645                     045646                     045647                     045648                     045649                     045650                     045651                     045652                     045653                     045654                     045655                     045656                     045657                     045658                     045659                     045660                     045661                     045662                     045663                     045664                     045665                     045666                     045667                     045668                     045669                     045670                     045671                     045672                     045673                     045674                     041391                     042686                     042687                     042688                     042689                     042690                     042691                     042692                     042693                     042694                     042695                     042696                     042697                     042698                     042699                     040631                     040632                     040633                     040634                     040635                     040636                     040637                     040638                     040639                     040640                     040641                     044308                     044309                     044310                     032020                     032021                     032022                     032023                     032024                     032025                     032026                     032027                     032028                     032029                     032030                     032031                     041975                     041976                     046289                     047171                     047172                     041969                     041970                     041971                     041972                     041973                     041974                     029880                     029881                     023848  025882  025883  025884  025885  029544                     029545                     029546                     029547                     030955  030956                     030960                     029879                     040504                     040505                     037779                     037780                     037781                     037782                     031928                     031929                     031930                     031931                     031932                     031933                     031934                     031935                     031936                     031937                     031938                     031939                     031940                     031941                     031942                     031943                     031944                     032055                     032056                     032057                     032058                     032059                     032060                     032061                     032062                     032063                     032064                     032065                     032066                     032067                     032068                     032069                     032070                     032071                     032072                     032073                     032074                     032075                     032076                     032077                     032078                     032079                     032080                     032081                     032082                     032083                     032084                     032085                     032086                     032087                     032088                     032089                     032090                     032091                     032092                     032093                     032094                     032095                     032096                     032097                     032098                     032099                     032100                     032101                     032102                     032103                     032104                     032105                     032106                     032107                     032108                     032109                     032110                     032111                     032112                     032113                     032114                     032115                     032116                     032117                     032118                     032119                     032120                     032121                     032122                     032123                     032124                     032125                     032126                     032127                     032128                     032129                     022843  022844  022845  024377  025716  045542                     045543                     045544                     030266  030485  042596                     042597                     042598                     042599                     042601                     042603                     042604                     026192  026510  026511  026516  026517  026518  034095  034096  034097  034098  034099  034517  035212  035213  035214  035215  035216  035217  035218  035219  035220  035221  035222  035223  035224  035225  035226  036935  036936  036937  036938  036939  036940  036941  037995  037996  037997  037998  037999  038000  038001  038002  038003  038004  038005  038006  038126  038127  038128  039790  039791  041998  042590  044486  045322  045910  045911  045912  045913  045914  045915  045916  047581  048338  048339  048340  048341  048342  048344  048345  048346  048347  048348  034424  034425  034518  034519  034520  034521  034522  034523  034524  034525  034526  034527  034528  034529  039359  039447  039448  039449  039450  039451  039452  041416  041417  041418  041419  041420  041421  041422  041423  041424  041425  041870  041871  041872  041873  041874  041875  041876  041877  041878  041879  041880  041881  041882  041883  041884  041885  042250  042251  042252  042253  042254  042255  042256  042257  042258  042259  044754  038909  039561  044713  044714  044715  044716  044717  044718  044719  044720  044721  044722  044723  044724  044725  044726  044727  044728  044729  044730  044731  044732  044733  044734  044735  044736  044737  044738  044739  044740  022903  022904  022905  034445  034446  035133  035134  035135  035136  035137  035138  035139  035140  035141  035142  035143  035144  035145  035146  035147  035148  035149  035150  038726  038727  038728  038729  038730  038731  038732  038733  038734  038735  038736  038737  038738  038739  038740  038741  038742  038743  038744  038745  038746  038747  038748  038749  038750  038751  038752  038753  027260  027261  027262  027263  027264  022224  022225  022226  022227  022228  022229  022231  022232  022233  022613  022614  022615  022616  022617  022618  022619  022620  022621  022622  022623  022624  022625  022626  022627  022628  022629  022630  022631  022632  022633  022634  022635  022636  022637  022638  022639  022640  022641  022642  022643  022644  022645  022646  022647  022648  022649  022650  022651  022652  022653  022654  022655  022656  022657  022658  022659  022660  022661  022662  022663  022664  022665  022666  022667  022668  022669  022670  022671  022672  022673  022674  022675  022676  022677  022678  022679  022680  022681  022682  022683  022684  022685  028406  028407  028408  028409  028410  028411  028412  028413  028414  028415  028416  028417  028418  028419  028420  028421  028422  028423  028424  028425  028426  028427  028428  028429  028430  028431  028432  028433  028434  028435  028436  028437  028438  028439  028440  028441  028442  028443  028444  028445  028446  028447  028448  028449  028450  028451  028452  028453  028454  028455  028456  028457  028458  028459  028460  028461  028462  028463  028464  028465  028466  028467  028468  028469  028470  028471  028472  028473  028474  028475  028476  028477  028478  028479  028480  028481  028482  028483  028484  028485  028486  022221  022222  022223  032470  032471  034616  034617  034618  034619  034620  041490  041658  041659  041660  041661  041686  041687  041688  041689  041690  041691  041692  041693  041694  041695  041696  041697  041698  041699  041700  041701  041702  041703  041704  041705  041706  041707  041708  041709  041710  041711  041712  041713  041714  041715  041716  041717  041718  041719  041720  041721  041722  041723  041724  041725  041726  041727  041728  041729  041730  041731  041732  041736  041737  041738  041739  041740  041741  041742  041743  041744  041745  041746  041747  041748  041749  041750  041751  041752  041753  041754  041755  041756  041757  041758  041759  041760  041761  041762  041763  041764  041765  041766  041767  041768  041769  041770  041771  041772  041773  041774  041775  041776  041777  041778  041779  041780  041781  041782  041783  041784  041785  041786  041787  041788  041789  041790  041791  041792  041793  041794  041795  041796  041797  041798  041799  041800  041801  041802  041803  041804  041805  041806  041807  041808  041809  041810  041811  041812  041813  041814  041815  041816  041817  041818  041819  041820  041821  041822  041823  041824  041825  041826  041827  041828  041829  041830  041831  041832  041833  041834  041835  041836  041837  041838  041839  041840  041841  041842  041843  041844  041845  041846  041847  041934  042591  042592  042593  044334  045325  045326  045327  045328  046794  046795  046796  046797  046798  047147  047582  047583  047713  047714  048498  048499  048500  048501  048502  048503  048504  022214  022215  022216  022217  022218  022219  022220  024303  026315  026316  026514  034421  041933  047897  047898  048041  048042  048043  048044  048045  048046  048047  048048  048049  048050  048051  048052  048053  048054  048055  048056  048057  048058  048059  048060  048061  048062  048063  048064  048065  048066  048067  048068  048069  048070  048071  048072  048073  048074  048075  048076  048077  048078  048079  048080  048081  048082  048083  048084  048085  048086  048087  048088  048089  048090  048091  048092  030970  0309",
      "center_classification_date": "20181109",
      "distribution_pattern": "The products were distributed to the following US states:  AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY.    The products were distributed to the following foreign countries:  Canada.",
      "state": "MI",
      "product_description": "The Spirit Plus Bed, Model 5600  Spirit Plus /with high rail siderail option",
      "report_date": "20181121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Medical Division of Stryker Corporation",
      "recall_number": "Z-0438-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81211",
      "termination_date": "20200428",
      "more_code_info": "71  030972  030973  030974  030975  030976  030977  030978  030979  030980  030981  030982  030983  030984  040881  040882  040883  040884  040885  040886  040887  040888  040889  040890  040891  040892  040893  040894  040895  040896  040897  040898  040899  040900  046458  047145  024359  024360  024361  024362  024363  024364  024365  024366  024367  024368  024369  024370  031128  031129  031130  031131  031132  031133  031134  031135  031136  031137  031138  031139  031140  031141  031142  031143  031144  031145  031146  031147  031148  031149  031150  031151  031152  031153  031154  031155  031156  031157  031158  031159  031160  031161  030296  030297  030298  030299  030300  030301  030302  030303  030304  030305  030306  030307  030308  030309  030310  030311  030312  030313  030314  030315  030316  030317  030318  030319  030320  030321  030322  030323  030324  030325  030326  030327  030328  030329  030330  030331  030332  030333  030334  030335  030336  030337  030338  030339  030340  030341  030342  030343  030344  030345  030346  030347  030348  030349  030350  030351  030352  030353  026193  027085  027086  027087  027088  027089  027090  027091  027092  027093  027094  027095  027096  027097  027098  027099  027100  027101  027102  027103  027104  027105  027106  027107  027108  027109  027110  027111  027112  027113  027114  027115  027116  027117  027118  027119  027120  027121  027122  027123  027124  027125  027126  027127  027128  027129  027130  027131  027132  027133  027134  027135  027136  027137  027138  027139  027140  027141  027142  027143  027144  027145  027146  027147  027148  027149  027150  027151  027152  027153  027154  027155  027156  027157  027158  027159  027160  027161  027162  027163  027164  027165  027166  027167  027168  027169  027170  027171  027172  027173  027174  027175  027176  027177  027178  027179  027180  027181  027182  027183  027184  027185  027186  027187  027188  027189  027190  027191  027192  027193  027194  027195  027196  027197  027198  027199  027200  027201  027202  027203  027204  027205  027206  027207  027208  027209  027210  027211  027212  027213  027214  028042  022264  022265  022266  022267  022268  022269  022270  022271  022272  022273  022274  022275  022276  022277  022278  022279  022280  022281  022282  022283  022284  022285  022286  022287  022288  022289  022290  022291  022292  022293  022294  022295  022296  022297  022298  022299  022300  022301  022302  022303  022304  022305  022306  022307  022308  022309  022310  022311  029878                     026501  026502  026503  026504  026505  026506  026507  026508  026509  026512",
      "recall_initiation_date": "20181008",
      "postal_code": "49002-5826",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}