{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81382",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of:  AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakistan, Portugal, Reunion, Saudi Arabia, Sweden, South Africa, Poland, Serbia, Russia, Turkey, Ukraine, China, India, Japan, Republic of Korea (South Korea), Malaysia, Thailand, and Taiwan.",
      "recall_number": "Z-0436-2019",
      "product_description": "QUADROX-i  Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120    The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.",
      "product_quantity": "490 total in US",
      "reason_for_recall": "There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4  x 3/16  that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.",
      "recall_initiation_date": "20181025",
      "center_classification_date": "20181128",
      "termination_date": "20200605",
      "report_date": "20181205",
      "code_info": "UDI: 04037691780139;  Lots: 92212343, 92214685, 92225187"
    }
  ]
}