      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86670",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:  Australia  Brazil  Canada  Chile  Colombia  Hong Kong  Italy  Mexico  Netherlands  Netherlands  New Zealand  South Africa  South Korea  Spain  Sweden  Switzerland  United Kingdom",
      "recall_number": "Z-0431-2021",
      "product_description": "Stryker Zyphr Disposable Cranial Perforator, Large 14/11 mm- is a sterile, single use cutting accessory intended for cutting an 11 mm diameter access hole through the cranium of adult patients.    Catalog Number: 5100-060-001",
      "product_quantity": "18,492  devices. Expanded Recall total: 18,594 devices",
      "reason_for_recall": "Potential for the inner bit to contain a crack, that may or may not be visible, which may lead to metal fragments entering the surgical site and/or delayed disengagment during use.",
      "recall_initiation_date": "20201016",
      "center_classification_date": "20201106",
      "termination_date": "20220622",
      "report_date": "20201118",
      "code_info": "GTIN: 04546540716224  Lot Numbers:  20029027  20030077  20030087  20031017  20031027  20034017  20035017  20035027  20035037  20036017  20036027  20038017  20038027  20040027  20041017  20042017  20044017  20044027  20047017  20048017  20048027  20048037  20051017  20052017  20052047  20055017  20055027  20055037  20055047  20056017  20057017  20057037  20058017  20058027  20059017  20059027  20059037  20059047  20062017  20062027  20062037  20062047  20063017  20063027  20064017  20066017  20066027  20066037  20066047  20069017  20069027  20071017  20071027  20072017  20079017  20079027  20080017  20083017  20083027  20084017  20084027  20084037  20084047  20084057  20094017  20094027  20094037  20094047  20099017  20099027  20099037  20101017  20101027  20101037  20101047  20101057  20101067  20101077  20106027  20106047  20106087  20106107  20106117  20106127  20106137  20107017  20107027  20107037  20107047  20107057  20108017  20108027  20108037  20108047  20108057  20108067  20111017  20111027  20111037  20112017  20112027  20112037  20112047  20113017  20113027  20113037  20118017  20118037  20120017  20120027  20120037  20121017  20121027  20122017  20122027  20122037  20122047  20125017  20127027  20127037  20127047  20128017  20128027  20128037  20128047  20128057  20129017  20131017  20131027  20131037  20134017  20134027  20136017  20136027  20136037  20136047  20136057  20136067  20136077  20139017  20139027  20139037  20141017  20141027  20141037  20141047  20142017  20142027  20142037  20142047  20143017  20143027  20146017  20146027  20147017  20147027  20147037  20147047  20148017  20148027  20148037  20150017  20150027  20150037  20150047  20154017  20154027  20155017  20155027  20155037  20156017  20157017  20157027  20157037  20161017  20162017  20162027  20164017  20167017  20167027  20168017  20168027  20168037    20171017  20171027  20171037  20171047  20171057  20171067  20171077  20171107  20177017  20177027  20177037  20178017  20178027  20178037  20189017  20189027  20189037  20189047  20189057  20189067  20190017  20190027  20192017  20195037  20197017  20199027  20199037  20199047  20199057  20199067  20202017  20202027  20203017  20203027  20203037  20204017  20204027  20204037  20204047  20206017  20206027  20206037  20206047  20206057  20206067  20211017  20211027  20211037  20211047  20217017  20217027  20217037  20220017  20220027  20220037  20220047  20220057    Expanded Recall 1/5/21:  20220067  20220077  20227027",