  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Allentown",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65356",
      "recalling_firm": "B. Braun Medical, Inc.",
      "address_1": "901 Marcon Boulevard",
      "address_2": "N/A",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-0424-2014",
      "product_description": "20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000    Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.",
      "product_quantity": "1,746,079 units",
      "reason_for_recall": "Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag.  Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution.  The misaligned cannula may result i",
      "recall_initiation_date": "20130404",
      "center_classification_date": "20131127",
      "termination_date": "20140826",
      "report_date": "20131204",
      "code_info": "Material BC2000, Lot numbers 0061275133, 0061275135, 0061275136, 0061279143, 0061279144, 0061279145, 0061279146, 0061289059, 0061289060, 0061289061, 0061289062, 0061289063, 0061289064, 0061289065, 0061289066, 0061289067 and 0061299664."
    }
  ]
}