  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "There is a potential for comingling of the 43MM and 51 MM molds",
      "address_2": "",
      "product_quantity": "72 units",
      "code_info": "Item Number 431207  Lot # 705550, UDI - (01)00880304447165(17)270927(10)705550    Item Number 431209   Lot # 705570, UDI - (01)00880304447141(17)270916(10)705570  Lot # 705580, UDI - (01)00880304447141(17)270916(10)705580",
      "center_classification_date": "20181107",
      "distribution_pattern": "US Nationwide Distribution in the states of:  AZ, FL, MI, OH, VA & WI",
      "state": "IN",
      "product_description": "StageOne Hip, Models 431207 & 431209Usage:      Product Usage:  The single-use cement spacer molds are sterile disposables made of medical grade silicone with a 316L stainless steel reinforcement stem. They are designed to be filled with polymethylmethacrylate/gentamicin bone cement, or equivalent, by injecting with a dispenser/gun into the mold. After the cement cures, the temporary spacers are to be removed from the molds with the reinforcement remaining as the core of the spacer, and placed into the joint space. The spacers remain in place (180 days or less) until the second stage of the two-stage procedure is performed to implant a conventional hip joint prosthesis.",
      "report_date": "20181114",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-0399-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81185",
      "termination_date": "20200414",
      "more_code_info": "",
      "recall_initiation_date": "20180921",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}