  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Louisville",
      "state": "CO",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89058",
      "recalling_firm": "Medtronic Navigation, Inc.",
      "address_1": "826 Coal Creek Cir",
      "address_2": "N/A",
      "postal_code": "80027-9710",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: MN, FL, MD, IL, AK, NE, NH, PA, ND, AR, CA, WI, NJ, GA, SD, AZ, KY, AL, MS, MO, TX, MI, MT, OH, NY, VA, MA, CT, TN, WV, WA, NC, CO, NV, ME, LA, DE, NM, IN, UT, DC, OR, SC, OK, IA, HI, ID, KS, WY, MN, PR  OUS: AU, BN, GU, JP, KR, MY, NZ, SG, TH, VN, MM, CA, CZ, RU, AL, CY, DE, ES, FR, GB, HU, IL, IT, NL, PL, RO, CN, HK, TW, IN, LK, NP, AR, BB, BR, CL, CO, CR, DO, EC, MX, PE, PR, VE, NI, UY, AE, BH, DZ, EG, IQ, JO, KZ, KE, SA, LB, LY, MA, NA, PK, QA, ZA, TR, AT, BE, CH, DK, FI, GF, GR, IE, LV, LT, LU, PT, SE, SK, UK, BA, BG, CI, HR, RS, SI",
      "recall_number": "Z-0383-2022",
      "product_description": "Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software.     Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585",
      "product_quantity": "2587",
      "reason_for_recall": "Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.",
      "recall_initiation_date": "20211111",
      "center_classification_date": "20211222",
      "report_date": "20211229",
      "code_info": "9733763, Version: 2.2.8;  9735585, Versions: 3.1.1, 3.1.2, 3.1.3"
    }
  ]
}