  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81159",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed to IN, NY, PR, and WV.",
      "recall_number": "Z-0382-2019",
      "product_description": "Liquid Cardiac Control, CQ5053",
      "product_quantity": "2108",
      "reason_for_recall": "Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.",
      "recall_initiation_date": "20181002",
      "center_classification_date": "20181102",
      "termination_date": "20190530",
      "report_date": "20181114",
      "code_info": "All lots",