  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Abbott Park",
      "address_1": "100 Abbott Park Rd",
      "reason_for_recall": "Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).",
      "address_2": "",
      "product_quantity": "26, 837 kits",
      "code_info": "Lot: 31497UN14 Exp.: 17 Oct. 2016; Lot: 09353UNI15 Exp.: 12 Feb. 2017; Lot: 45216UN15 12 May 2017; Lot: 14596UN15 Exp.: 30 Jul. 2017; Lot: 37055UN15 Exp.: 12 Oct. 2017; Lot: 14782UN16 Exp.: 21 Apr. 2018",
      "center_classification_date": "20161109",
      "distribution_pattern": "Nationwide and Worldwide Distribution.",
      "state": "IL",
      "product_description": "ARCHITECT Lactic Acid Part # 09D891T21    The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.",
      "report_date": "20161116",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Abbott Laboratories",
      "recall_number": "Z-0379-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75448",
      "termination_date": "20170517",
      "more_code_info": "",
      "recall_initiation_date": "20161011",
      "postal_code": "60064-3502",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}