  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72555",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to states of: CA, FL, MI  and WI.",
      "recall_number": "Z-0375-2016",
      "product_description": "Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System) The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures.  Intramedullary fixation rod, orthopedic.",
      "product_quantity": "5",
      "reason_for_recall": "TFNA nails may have the locking mechanism too close to the top of the nail. This may 1.  Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the bottom of the cannulated connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.",
      "recall_initiation_date": "20151102",
      "center_classification_date": "20151204",
      "termination_date": "20160513",
      "report_date": "20151216",
      "code_info": "Part Number Lot Number  04.037.162S 7950511",
      "more_code_info": ""
    }
  ]
}