{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93323",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA,  CANADA, CHILE, JAPAN, NETHERLANDS, NEW ZEALAND.",
      "recall_number": "Z-0374-2024",
      "product_description": "Compress Device Segmental Anchor Plug, 18 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.  Item Number: 178408",
      "product_quantity": "N/A",
      "reason_for_recall": "Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery",
      "recall_initiation_date": "20231009",
      "center_classification_date": "20231121",
      "report_date": "20231129",
      "code_info": "Lot Numbers/UDI: 017910    (01)00880304259768(17)270807(10)017910; 017910R  (01)00880304259768(17)270807(10)017910R; 018060    (01)00880304259768(17)270817(10)018060; 332060    (01)00880304259768(17)270817(10)332060; 608610    (01)00880304259768(17)271204(10)608610; 671740    (01)00880304259768(17)271117(10)671740; 671740R   (01)00880304259768(17)271117(10)671740R; 752790    (01)00880304259768(17)270816(10)752790; 752790R  (01)00880304259768(17)270816(10)752790R; 965290    (01)00880304259768(17)271030(10)965290;",
      "more_code_info": ""
    }
  ]
}