  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93282",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0352-2024",
      "product_description": "Tina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Material Numbers 03001245322 - (cobas c 311, 501 and 502; INTEGRA 400 plus), 08105618190 - (cobas c 303 and 503)",
      "product_quantity": "5,263 kits",
      "reason_for_recall": "Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.",
      "recall_initiation_date": "20231003",
      "center_classification_date": "20231120",
      "report_date": "20231129",
      "code_info": "All Lots Affected.  UDI: 04015630914593 - (cobas c 311, 501 and 502; INTEGRA 400 plus) UDI: 07613336133651 - (cobas c 303 and 503)",
      "more_code_info": ""
    }
  ]
}