  "results": [
    {
      "status": "Terminated",
      "city": "Kansas City",
      "state": "MO",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91054",
      "recalling_firm": "Primus Corporation dba Trinity Biotech",
      "address_1": "4231 E 75th Ter",
      "address_2": "",
      "postal_code": "64132-2059",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.",
      "recall_number": "Z-0349-2023",
      "product_description": "A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis",
      "product_quantity": "340 kits",
      "reason_for_recall": "The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months.  The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.",
      "recall_initiation_date": "20211005",
      "center_classification_date": "20221201",
      "termination_date": "20230809",
      "report_date": "20221207",
      "code_info": "Lot 12161",
      "more_code_info": ""
    }
  ]
}