  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lenoir",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91078",
      "recalling_firm": "Greer Laboratories, Inc.",
      "address_1": "639 Nuway Cir",
      "address_2": "",
      "postal_code": "28645-3646",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Distribution to states of:  CA, CO, IA, KY, MA, MD, MO, NC, PA, UT, VA, and International to countries of: Canada, Denmark",
      "recall_number": "Z-0347-2023",
      "product_description": "GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.",
      "product_quantity": "218 units",
      "reason_for_recall": "The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency.  The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.",
      "recall_initiation_date": "20221025",
      "center_classification_date": "20221130",
      "termination_date": "20260211",
      "report_date": "20221207",
      "code_info": "Lot Number 389591, exp 6/12/2023",
      "more_code_info": ""
    }
  ]
}