  "results": [
    {
      "status": "Ongoing",
      "city": "Aizuwakamatsu",
      "state": "N/A",
      "country": "Japan",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93222",
      "recalling_firm": "Aizu Olympus Co., Ltd.",
      "address_1": "3 Chome 1-1",
      "address_2": "Niiderakita",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-0340-2024",
      "product_description": "Mobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management.  Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.",
      "product_quantity": "33 units",
      "reason_for_recall": "Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection",
      "recall_initiation_date": "20230925",
      "center_classification_date": "20231117",
      "report_date": "20231129",
      "code_info": "Model Numbers: MAF-DM2 (04953170407260), MAF-GM (04953170340369), MAF-GM2 (04953170407208), MAF-TM (04953170288630), MAF-TM2 (04953170407147).",
      "more_code_info": ""
    }
  ]
}