  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86521",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0333-2021",
      "product_description": "ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22mm, Neutral, O.D. Cup with Spacers, 51 mm    Item Number: 00-8065-548-22",
      "product_quantity": "412 unit total",
      "reason_for_recall": "Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant.  Cannot rule out the potential for adverse tissue reactions.",
      "recall_initiation_date": "20200924",
      "center_classification_date": "20201102",
      "termination_date": "20220520",
      "report_date": "20201111",
      "code_info": "All lots"
    }
  ]
}