  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83962",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed nationwide.",
      "recall_number": "Z-0320-2020",
      "product_description": "Fitmore Hip Stem, hip prosthesis, various sizes, Item Nos. 0100551102  0100551103  0100551104  0100551105  0100551106  0100551107  0100551108  0100551109  0100551110  0100551111  0100551112  0100551113  0100551114  0100551201  0100551202  0100551203  0100551204  0100551205  0100551206  0100551207  0100551208  0100551209  0100551210  0100551211  0100551212  0100551213  0100551214  0100551301  0100551302  0100551303  0100551304  0100551305  0100551306  0100551307  0100551308  0100551309  0100551310  0100551311  0100551312  0100551313  0100551314  0100551401  0100551402  0100551403  0100551404  0100551405  0100551406  0100551407  0100551408  0100551409  0100551410  0100551411  0100551412  0100551413  0100551414",
      "product_quantity": "5,459,583 total devices",
      "reason_for_recall": "Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).",
      "recall_initiation_date": "20191010",
      "center_classification_date": "20191107",
      "termination_date": "20200519",
      "report_date": "20191113",
      "code_info": "All product manufactured prior to January 2014."
    }
  ]
}