  "results": [
    {
      "status": "Ongoing",
      "city": "Bowdoin",
      "state": "ME",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95479",
      "recalling_firm": "FHC, Inc.",
      "address_1": "1201 Main St",
      "address_2": "N/A",
      "postal_code": "04287-7302",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution only to the following states: AZ, CA, PA, TN, TX, WA.",
      "recall_number": "Z-0312-2025",
      "product_description": "microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide a microelectrode or instrument in the brain during stereotactic functional neurosurgical procedures.",
      "product_quantity": "8 packs of 5ea (40 total)",
      "reason_for_recall": "A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.",
      "recall_initiation_date": "20241004",
      "center_classification_date": "20241108",
      "report_date": "20241120",
      "code_info": "Package DI: 10873263004323; Contains DI Package: 00873263004326 LOT No. 244517"
    }
  ]
}