{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81403",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0305-2019",
      "product_description": "Brilliance iCT (Model No. 728306), Software Version 4.1.6    Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.",
      "product_quantity": "626",
      "reason_for_recall": "It was reported that scanners running certain software versions provide confusing information in the \"Impressions\"  section of the Calcium Score report. The generated report provides multiple  analyses of the amount of plaque detected, as if they all apply to the patient at  the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.",
      "recall_initiation_date": "20170607",
      "center_classification_date": "20181031",
      "termination_date": "20190703",
      "report_date": "20181107",
      "code_info": "Software Version 4.1.6"
    }
  ]
}