{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "345 E Main St",
      "reason_for_recall": "Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank.  Based on  investigation, this problem appears to be due to a counterfeit chip on the",
      "address_2": "",
      "product_quantity": "204",
      "code_info": "Catalog 60-3000-101-00,   Serial # 3005IACJ, 3006JAEP, 3007CADR, 3007EACT,  3007EAKT, 3007GAGR, 3007HABK, 3007KAAR,  3007KAET, 3008CADE, 3008HAHC, 3008LACK,  3009AACN, 3009HABA, 3009LAEF, 3010CABH,  3010DAHG, 3010GAAC, 3010GAAJ, 3011LABL,  3011LABN, 3011LABQ, 3011LACC, 3011LACE,  3011LACF, 3011LACG, 3011LACH, 3011LACJ,  301 1LACK, 3011LACL, 3011LACM, 3011LACN,  3011LACP, 3011LACQ, 3011LACR, 3011LACS,  3011LACT, 3011LACU, 3011LACW, 3011LADA,  3011LADE, 3011LADF, 3011LADG, 3011LADH,  3011LADJ, 3011LADK, 3011LADM, 3011LADP,  3011LADQ, 3011LADR, 3011LADS, 3011LADT,  3011LADU, 3011LADW, 3011LAEA, 3011LAEB,  3011LAEC, 3011LAED, 3011LAEE, 3011LAEF,  3011LAEH, 3012AAAG, 3012AAAK, 3012AAAM,  3012AAAN, 3012AAAP, 3012AAAQ, 3012AAAR,  3012AAAS, 3012AAAT, 3012AAAU, 3012AAAW,  3012AABA, 3012AABB, 3012AABC, 3012AABE,  3012AABG, 3012AABH, 3012AABK, 3012AABL,  3012AABM, 3012AABN, 3012AABQ, 3012AABR,  3012AABS, 3012AABU, 3012AACA, 3012AACB,  3012AACC, 3012AACD, 3012AACE, 3012AACH,  3012AACJ, 3012AACM, 3012AACN, 3012AACP,  3012AACQ, 3012AACR, 3012AACS, 3012AACT,  3012AACW, 3012AADA, 3012AADB, 3012AADC,  3012AADD, 3012AADE, 3012AADF, 3012AADG,  3012BAAA, 3012BAAB, 3012BAAC, 3012BAAD,  3012BAAE, 3012BAAF, 3012BAAG, 3012BAAH,  3012BAAJ, 3012BAAK, 3012BAAL, 3012BAAM,  3012BAAN, 3012BAAP, 3012BAAQ, 3012BAAR,  3012BAAS, 3012BAAT, 3012BAAU, 3012BAAW,  3012BABA, 3012BABB, 3012BABC, 3012BABD,  3012BABE, 3012BABF, 3012BABG, 3012BABH,  3012BABJ, 3012BABK, 3012BABL, 3012BABM,  3012BABN, 3012BABP, 3012BABQ, 3012BABR,  3012BABS, 3012BABT, 3012BABU, 3012BABW,  3012BACE, 3012BACG, 3012BACH, 3012BACJ,  3012BACK, 3012BACL, 3012BACM, 3012BACN,  3012BACP, 3012BACQ, 3012BACR, 3012BACS,  3012BACT, 3012BACU, 3012BACW, 3012BADA,  3012BADB, 3012BADC, 3012BADD, 3012BADG,  3012BADH, 3012BADJ, 3012BADK, 3012BADL,  3012BADN, 3012BADP, 3012BADR, 3012BADS,  3012BADU, 3012BADW, 3012BAEA, 3012BAEB,  3012BAEC, 3012BAEH, 3012BAEJ, 3012BAEL,  3012BAEM, 3012BAEN, 3012BAEQ, 3012BAER    French Catalog 60-3000 301-02 Serial # 3012 AAAE  German Catalog 60-300060-301-03, Serial # 3012AAAF  Spanish Catalog 60-3000-301-11, Serial # 3012AAAB, 3012AAAC, 3012AACG, 3012BACA, 3012BACB, 3012BACC, 3012BACD  Dutch 60-3000-301-12, Serial # 3012AAAA, 3012AAAD, 3012EACA",
      "center_classification_date": "20121115",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Belgium, Czech Republic, Egypt, France, Germany, Morocco, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey, and UAE.",
      "state": "IN",
      "product_description": "ATS 3000 Automatic Tourniquet System  w/HOSES AND LOP SENSOR      Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.",
      "report_date": "20121121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-0296-2013",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "63458",
      "termination_date": "20140417",
      "more_code_info": "",
      "recall_initiation_date": "20120905",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}