{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Haifa",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97647",
      "recalling_firm": "GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING",
      "address_1": "9, Andrey Sakharov",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-0295-2026",
      "product_description": "Elscint Helix; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging",
      "product_quantity": "3 units",
      "reason_for_recall": "GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.",
      "recall_initiation_date": "20250912",
      "center_classification_date": "20251023",
      "report_date": "20251029",
      "code_info": "all serial numbers in distribution",
      "more_code_info": ""
    }
  ]
}