  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "North Billerica",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81074",
      "recalling_firm": "Magellan Diagnostics, Inc.",
      "address_1": "101 Billerica Ave Bldg 4",
      "address_2": "N/A",
      "postal_code": "01862-1271",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US (Nationwide) and countries of: Spain and South Africa.",
      "recall_number": "Z-0285-2019",
      "product_description": "Magellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood.  Catalog Number: 70-6762  Primary DI No:  00850355006000     The LeadCare II Blood Lead Test Kit is for in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and treatment reagent of the test kit are specific for lead only. Contents of the LeadCare II Blood Lead Test Kit must only be used with a LeadCare II Analyzer.",
      "product_quantity": "1,465 kits",
      "reason_for_recall": "Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 ¿g/dL). Results for patient testing with this lot will be displayed as HIGH.",
      "recall_initiation_date": "20180808",
      "center_classification_date": "20181026",
      "termination_date": "20200512",
      "report_date": "20181107",
      "code_info": "Lot 1808M and sub-lots: 1808M-01, 1808M-02, 1808M-03, 1808M-04, 1808M-05, 1808M-06, 1808M-07, 1808M-08, 1808M-09.    Exp. Date:November 10, 2019"
    }
  ]
}