  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Boardman",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63083",
      "recalling_firm": "Primus Medical LLC",
      "address_1": "8401 Southern Blvd",
      "address_2": "N/A",
      "postal_code": "44512-6709",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution: USA including states of: AZ, IL, KY, MD, MI, OH, OR, PA, TX and VA.",
      "recall_number": "Z-0270-2013",
      "product_description": "AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512.    Usage: Hospital Bed.",
      "product_quantity": "342 Beds",
      "reason_for_recall": "The \"Retaining washers & Clevis Pins\" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.",
      "recall_initiation_date": "20110202",
      "center_classification_date": "20121108",
      "termination_date": "20140514",
      "report_date": "20121114",
      "code_info": "Model #TL500; Serial Number: 001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024, 025, 026, 027, 028, 029, 030, 031, 032, 033, 034, 035, 039, 040, 041, 042, 043, 044, 045, 046, 047, 048, 049, 050, 052, 053, 054, 055, 056, 057, 058, 059, 060, 061, 062, 063, 064, 065, 070, 075, 081, 09350014, 09350015, 09350016, 09350017, 09350018, 09350021, 09350023, 09350025, 09350026, 09350027, 09350028, 09350029, 09350030, 09350031, 09350032, 09350033, 09350035, 09350038, 09350040, 09350041, 09350042, 09350045, 09350048, 09350049, 09350054, 09350055, 09350056, 09350057, 09350058, 09350059, 09350060, 09350062, 09350063, 09350064, 09350065, 09350066, 09350067, 09350068, 09350069, 09350070, 09350072, 09350074, 09350082, 09350085, 09350087, 09350088, 09350089, 09350090, 09350091, 09350092, 09430001, 09430002, 09430003, 09430004, 09430005, 09430006, 09430007, 09430008, 09430010, 09430013, 09430014, 09430015, 09430017, 09430019, 09430020, 09430021, 09430022, 09430024, 09430026, 09430027, 09430028, 09430029, 09430030, 09430032, 09430033, 09430034, 09430036, 09430037, 09430038, 09430039, 09430040, 09430041, 09430042, 09430043, 09430044, 09430045, 09430046, 09430047, 09430049, 09430050, 09430051, 09430052, 09430053, 09430054, 09430056, 09430057, 09430058, 09430059, 09430060, 09430061, 09430062, 09430063, 09430065, 09430066, 09463067, 09430068, 09430069, 09430070, 09430071, 09430072, 09430074, 09430075, 09430076, 09430077, 09430078, 09430079, 09430080, 09430081, 09430083, 09430085, 09430086, 09430087, 09430088, 09430089, 09430091, 09430093, 09430094, 09430097, 09430099, 09430100, 09520001, 09520002, 09520003, 09520004, 09520005, 09520006, 09520007, 09520008, 09520009, 09520011, 09520012, 09520013, 09520015, 09520017, 09520018, 09520019, 09520020, 09520021, 09520022, 09520025, 09520026, 09520027, 09520028, 09520029, 09520030, 09520031, 09520032, 09520033, 09520034, 09520035, 09520036, 09520037, 09520038, 09520039, 09520040, 09520041, 09520042, 09520043, 09520044, 09520045, 09520046, 09520047, 09520048, 09520049, 09520050, 10040001, 10040003, 10040004, 10040005, 10040006, 10040007, 10040008, 10040009, 10040010, 10040011, 10040012, 10040013, 10040014, 10040015, 10040016, 10040017, 10040028, 10040029, 10040030, 10040032, 10040033, 10040034, 10040035, 10040036, 10040037, 10040038, 10040039, 10040040, 10040041, 10040042, 10040043, 10040044, 10040045, 10040046, 10040047, 10040048, 10040049, 10040050, 10040051, 10040052, 10040053, 10040054, 10040055, 10040056, 10040057, 10040058, 10040059, 10040062, 10040063, 10040064, 10040065, 10040066, 10040067, 10040068, 10040069, 10040070, 10040071, 10040073, 10040074, 10040075, 10040076, 10040077, 10040078, 10400029, 10400050, 10400056, 10400057, 10400058, 10400059, 10400065, 10370016, 10370017, 10370018, 10370019, 10370020, 10370032, 10370033, 10370034, 10370035, 10370038, 10370052, 10370053, 10370054, 10370056, 10370060, 10370063, 10370078, 10370079, 10370080 , 09380031, 09380029, 09380064, 09380070, 09380089, 09380040, 09380038, 09380085, 09380045, 09380054, 09380060, 09380059, 09380058, 09380048, 09380042, 09380038, 09380069, 09380063, 09380067 & 09380065."
    }
  ]
}