  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "South Portland",
      "state": "ME",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99956",
      "recalling_firm": "Solstice Corporation",
      "address_1": "85 E St",
      "address_2": "",
      "postal_code": "04106-2870",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states and territory of CA, FL, IL, IN, MN, MO, NY, OH, PA, TX, WV and Puerto Rico. The country of Canada.",
      "recall_number": "Z-0268-2027",
      "product_description": "Xact mole marker for mammography  Model/Catalog Number: M15RM",
      "product_quantity": "736",
      "reason_for_recall": "The radiopaque material may detach from the marker. In some cases, artifact was noted and attributed to the detached material.",
      "recall_initiation_date": "20260922",
      "center_classification_date": "20261022",
      "report_date": "20261028",
      "code_info": "part #: M15RM; UDI-DI: 00854623006068;  UPC-A; lot #: M15RM26A;",
      "more_code_info": ""
    }
  ]
}