{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99949",
      "recalling_firm": "Medtronic Neuromodulation",
      "address_1": "7000 Central Ave Ne",
      "address_2": "",
      "postal_code": "55432-3568",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and territory of Puerto Rico. The countries of Brazil, Colombia, Chile, Canada, Japan, New Zealand, Australia, United Kingdom, Switzerland, Sweden, Spain, Portugal, Italy, Israel, Germany, France, Finland, Canary Islands, Belgium, Austria, Norway, Northern Ireland, Netherlands, Iceland, Denmark, South Africa, Turkey, Slovakia, Saudi Arabia, Poland, Hungary, United Arab Emirates (UAE), Kuwait, Greece.",
      "recall_number": "Z-0267-2027",
      "product_description": "InterStim Percutaneous Extensions     EXT 3560022 ISTM II PERC\t  \t3560022  \t  EXT 3560030 ISTM MICRO PERC\t  \t3560030",
      "product_quantity": "56,331 (US); 40,211 (OUS)",
      "reason_for_recall": "Wrong lead or percutaneous extension selected. Use of an incompatible lead or percutaneous extension will prevent the system from working as intended and may result in delay of surgery or need for surgical replacement of the product.",
      "recall_initiation_date": "20260908",
      "center_classification_date": "20261022",
      "report_date": "20261028",
      "code_info": "3560022    UDI-DI 00763000203894    Lots VA24HE7   VA24R5J   VA24TXT   VA24UYB   VA24WJE   VA24WJF   VA250XH   VA25144   VA2515A   VA25TEF   VA25TEH   VA25TEJ   VA25TEL   VA26A7F   VA26AJA   VA26AJB   VA26AKS   VA26AKT   VA26AKU   VA26AKV   VA26C4T   VA26CB1   VA26CB2   VA26EHL   VA26EHN   VA26EHP   VA2725R   VA2725S   VA2726H   VA2726J   VA2728H   VA2728K   VA2728L   VA275QE   VA2760F   VA277W0   VA277W1   VA27C85   VA27C86   VA27CCG   VA27CEE   VA27D4B   VA27D62   VA27D63   VA27KEH   VA27KGX   VA27KHG   VA27KHH   VA27KHJ   VA27KHK   VA27KHL   VA27NS0   VA27NSC   VA27NSG   VA27NSH   VA27NSJ   VA27NSL   VA27ST1   VA27ST2   VA27UGK   VA27UGL   VA27UXN   VA27UXP   VA27VDZ   VA27VED   VA27VEE   VA27VEF   VA27VJF   VA27VJG   VA27VJJ   VA27VWY   VA27Y21   VA27Y22   VA27YF8   VA27YFG   VA280D1   VA280E6   VA280LQ   VA2848G   VA28499   VA284MR   VA284PJ   VA284S3   VA286PS   VA286PT   VA2872W   VA28AQH   VA28BDU   VA28BEW   VA28BR6   VA28BR7   VA28DZR   VA28E02   VA28EC2   VA28EC3   VA28GE3   VA28HC2   VA28HM9   VA28HSG   VA28J1R   VA28J1S   VA28K8K   VA28MHH   VA28MHS   VA28MYN   VA28N0Y   VA28N15   VA28N1R   VA28T7D   VA28T7F   VA28TSM   VA28TSP   VA28UE7   VA28UHR   VA28UHY   VA28Z99   VA28ZJ5   VA28ZMP   VA2976F   VA297CK   VA297CR   VA297PV   VA2990L   VA29939   VA299J7   VA299J8   VA29EQE   VA29EQF   VA29EQG   VA29EQJ   VA29FSC   VA29FSD   VA29FSS   VA29G5U   VA29G94   VA29GKR   VA29HCQ   VA29HCU   VA29HM5   VA29HM6   VA29N97   VA29QAK   VA29QAL   VA29QBK   VA29QBL   VA29QCE   VA29QUW   VA29QUZ   VA29QVN   VA29S6N   VA29S6Q   VA29SAV   VA29SDN   VA29SEY   VA29SEZ   VA29SF4   VA29SF5   VA29SYJ   VA29Y62   VA29Y63   VA29YGH   VA29YGJ   VA29YKY   VA29YW5   VA29Z8P   VA29Z8Q   VA29ZEG   VA29ZN3   VA29ZSA   VA29ZU3   VA29ZU4   VA29ZUG   VA2A2Z3   VA2A5JQ   VA2A5MK   VA2A6GU   VA2A6GY   VA2A6L5   VA2A73R   VA2A73U   VA2A7EC   VA2A7EF   VA2A7HW   VA2AEKR   VA2AEM9   VA2AEWP   VA2ALQ9   VA2ALUD   VA2ALV0   VA2ALVA   VA2ALW5   VA2APNN   VA2APQA   VA2APSH   VA2APSU   VA2ARAX   VA2ARCA   VA2ARCT   VA2AYRR   VA2AYRU   VA2AYZ3   VA2AZ0W   VA2AZ7J   VA2AZ7R   VA2B6PX   VA2B8WD   VA2B8WE   VA2B8XZ   VA2B8ZN   VA2BBDN   VA2BBDQ   VA2BBT6   VA2BBU0   VA2BBWC   VA2BBZM   VA2BD21   VA2BDCD   VA2BDCE   VA2BHB0   VA2BHB9   VA2BHCD   VA2BHCF   VA2BHDU   VA2BHDX   VA2BHJN   VA2BHPH   VA2BHPK   VA2BHPR   VA2BUNB   VA2BUNP   VA2BUSW   VA2BUVT   VA2BXXB   VA2BY1R   VA2BY2N   VA2BY4M   VA2C4A9   VA2C4JZ   VA2C582   VA2C585   VA2C7JX   VA2C7LT   VA2C7SB   VA2CCJ0   VA2CCJ1   VA2CCJK   VA2CCJL   VA2CCJM   VA2CCJN   VA2CCJP   VA2CCJR   VA2CH8Y   VA2CHC0   VA2CHET   VA2CM6R   VA2CM9W   VA2CMA3   VA2CMAG   VA2CMAL   VA2CN36   VA2CPWP   VA2CQ0U   VA2CQ0Z   VA2CQ1G   VA2CQ23   VA2CQ2A   VA2CSVR   VA2CSW9   VA2CSWG   VA2CSX2   VA2CT5M   VA2CUN8   VA2CVB8   VA2CW13   VA2CZSU   VA2CZSV   VA2CZSY   VA2CZT4   VA2D4PR   VA2D4Q5   VA2D4QJ   VA2D4QM   VA2D4Z4   VA2D4ZU   VA2D657   VA2D658   VA2D659   VA2D65A   VA2D65B   VA2D8B5   VA2DDCK   VA2DDCU   VA2DDD8   VA2DDDG   VA2DDDT   VA2DP4Z   VA2DPAJ   VA2DPCV   VA2DPF5   VA2DPNR   VA2DVAY   VA2DVB5   VA2DVBA   VA2DVBF   VA2DVDA   VA2DVP5   VA2DVPQ   VA2E1PW   VA2E1PY   VA2E1Q1   VA2E1QE   VA2E2EK   VA2E491   VA2E494   VA2E4A5   VA2E4AC   VA2ECQY   VA2ECRM   VA2ECS4   VA2ECTK   VA2ECTM   VA2ECTV   VA2EGXE   VA2EGXJ   VA2EGXK   VA2EGZ4   VA2EHT8   VA2EMDT   VA2EME1   VA2EME9   VA2EMEP   VA2EMF5   VA2EMFB   VA2EMM1   VA2EMN8   VA2ENF9   VA2ENH4   VA2ENH7   VA2ENJG   VA2ENJQ   VA2ENJT   VA2ENJV   VA2ENKX   VA2ENW3   VA2EP8C   VA2ESRT   VA2ESSH   VA2ESSW   VA2ESTM   VA2ESTV   VA2ESU0   VA2ESUA   VA2ESUK   VA2EYMM   VA2EYNF   VA2EYQQ   VA2EYRB   VA2F0TM   VA2F0U6   VA2F0UV   VA2F1B5   VA2F1G2   VA2F1G6   VA2F1GQ   VA2F1HH   VA2F1XE   VA2F1XY   VA2F1ZF   VA2F202   VA2F21Q   VA2F222   VA2F259   VA2F25E   VA2F25V   VA2F38C   VA2F38K   VA2F38T   VA2F3DE   VA2F3ET   VA2F3FG   VA2F3FX   VA2F3G6   VA2F3K6   VA2F4BJ   VA2F4E5   VA2F4KD   VA2F4M7   VA2F4MY   VA2FBYT   VA2FBYW   VA2FBZJ   VA2FBZZ   VA2FC00   VA2FC0H   VA2FC0T   VA2FPRU   VA2FWNY   VA2FWP1   VA2FWPK   VA2FWQJ   VA2FZRU   VA2FZRY   VA2FZRZ   VA2FZS1   VA2FZS8   VA2FZSB   VA2G05M   VA2G05U   VA2G06S   VA2G06W   VA2G06Z   VA2G074   VA2G07T   VA2G4WX   VA2G4X0   VA2G4X5   VA2G4XS   VA2G4XV   VA2G6MT   VA2G6NL   VA2G6NZ   VA2G6R9   VA2G6RE   VA2G79V   VA2G7A4   VA2G8J5   VA2G8J7   VA2G8J8   VA2G8S6   VA2G8UC   VA2G8UH   VA2G8WC   VA2G8WP   VA2G8WV   VA2G8X1   VA2G8X9   VA2G8XG   VA2G8YN   VA2G8Z6   VA2GDZE   VA2GDZF   VA2GDZH   VA2GDZJ   VA2GDZK   VA2GDZM   VA2GEA9   VA2GEAV   VA2GEBH   VA2GEDS   VA2GEEP   VA2GEF3   VA2GH1M   VA2GH1W   VA2GH25   VA2GH2F   VA2GH3Q   VA2GH9H   VA2GKGQ   VA2GKGW   VA2GKH1   VA2GKH6   VA2GL2M   VA2GL3F   VA2GL3N   VA2GL6Z   VA2GT1C   VA2GT1J   VA2HF33   VA2HF3H   VA2HF3Y   VA2HJJU   VA2HJJX   VA2HJK5   VA2HJX0   VA2HJX4   VA2HJXN   VA2HK92   VA2HK98   VA2HK9B   VA2HK9C   VA2HKK0   VA2HKKE   VA2HLVK   VA2HLVQ   VA2HLW7   VA2HLW9   VA2HLWC   VA2HLWD   VA2HLWE   VA2HMQR   VA2HMR1   VA2HMRS   VA2HMS7   VA2HMSN   VA2HMSQ   VA2HMTK   VA2HT5K   VA2HTGH   VA2HTGU   VA2HTJ6   VA2HTJP   VA2HTK8   VA2HTKL   VA2HTLR   VA2HTP3   VA2HTPB   VA2J72W   VA2J73J   VA2J74K   VA2J74R   VA2J7C2   VA2J7D5   VA2J7KQ   VA2J7L8   VA2J7LN   VA2J7ME   VA2J7ND   VA2J7VP   VA2J7VV   VA2J7VZ   VA2J7WD   VA2JL82   VA2JL87   VA2JL8P   VA2JLBK   VA2JLBU   VA2JLUV   VA2JLXK   VA2JLXM   VA2JLYG   VA2JLZS   VA2JLZT   VA2JLZU   VA2JLZV   VA2JLZW   VA2JLZX   VA2JMW2   VA2JMWK   VA2JMXM   VA2JMYA   VA2JMYG   VA2JMZJ   VA2JN1B   VA2JN1M   VA2JN2V   VA2JQF8   VA2JQG0   VA2JQG5   VA2JQGJ   VA2JQGQ   VA2JQZG   VA2JQZP   VA2JQZY   VA2JR04   VA2JR0F   VA2JR0J   VA2JR7R   VA2JR86   VA2JR8A   VA2JR8D   VA2JR8L   VA2JR8M   VA2JRVJ   VA2JRXU   VA2JRYA   VA2JRYE   VA2JRYZ   VA2JS0M   VA2JSSZ   VA2JST1   VA2JSTF   VA2JSTH   VA2JT5P   VA2JT5Q   VA2JT5Z   VA2JT60   VA2JTCE   VA2JTDT   VA2JWBU   VA2JWCA   VA2JWDV   VA2JWE5   VA2JWEC   VA2JWF4   VA2JWT7   VA2JWTU   VA2JWU8   VA2JWVF   VA2JWVT   VA2JWWC   VA2JWXG   VA2JX2S   VA2JX2Z   VA2JX3E   VA2K89B   VA2K8A0   VA2K8AA   VA2KUVM   VA2KUWM   VA2KUWT   VA2KVCU   VA2KVD5   VA2KVDC   VA2KVP4   VA2KVQU   VA2KVQY   VA2KVRA   VA2KVU4   VA2KVX9   VA2KVXC   VA2KW4P   VA2KW4R   VA2KWQW   VA2KWSG   VA2KWSJ   VA2KWSP   VA2KWSU   VA2KWSY   VA2KWTS   VA2KWU1   VA2KWUE   VA2KWUZ   VA2KX2P   VA2KX51   VA2KX5D   VA2KX5N   VA2KX61   VA2KX6F   VA2L1YH   VA2L1YK   VA2L1YM   VA2L263   VA2L26M   VA2L2C7   VA2L2DP   VA2L2E0   VA2L7J0   VA2L7JE   VA2L7KW   VA2L88B   VA2L88N   VA2L892   VA2L898   VA2L89K   VA2L9ZG   VA2L9ZK   VA2LA10   VA2LA1B   VA2LA1F   VA2LA1J   VA2LA8G   VA2LA9N   VA2LAA8   VA2LABL   VA2LAC8   VA2LACS   VA2LH21   VA2LH35   VA2LH4C   VA2LH4S   VA2LH63   VA2LH6F   VA2LH6G   VA2LH6L   VA2LTAZ   VA2LTC1   VA2LTE6   VA2LTEQ   VA2LU6K   VA2LU6P   VA2LUA5   VA2LUCH   VA2LUES   VA2LUFW   VA2LUJ4   VA2LUJG   VA2LVG8   VA2LVGG   VA2LVGR   VA2LVLA   VA2LVT2   VA2LVTM   VA2LVTV   VA2LVUK   VA2LVX2   VA2LVXH   VA2LWED   VA2LWEE   VA2LWEH   VA2LWM7   VA2M0GB   VA2M0GK   VA2MHSZ   VA2N7R7   VA2N7S3   VA2N7TZ   VA2N7W1   VA2N7W4   VA2N7Z5   VA2N7ZQ   VA2N7ZY   VA2N80G   VA2N81E   VA2N8GG   VA2N8HQ   VA2N8RZ   VA2N8TM   VA2NCAS   VA2NCB1   VA2NCDL   VA2NCE0   VA2NCF7   VA2NCGR   VA2NCJ2   VA2NCLH   VA2NCLR   VA2NCLX   VA2NCMW   VA2NCPG   VA2NCR3   VA2NCRG   VA2NCS5   VA2NCS7   VA2NN6S   VA2NN75   VA2NN78   VA2NN7R   VA2NN7W   VA2NN95   VA2NN96   VA2NN9V   VA2NNP5   VA2NNP8   VA2NNPP   VA2NNQ8   VA2NNRP   VA2NNSP   VA2NNUW   VA2NP1Q   VA2NP2B   VA2NP2H   VA2NP2K   VA2NP2Z   VA2NWP5   VA2NWQ7   VA2NWQN   VA2NWRH   VA2NWRY   VA2NWS1   VA2NWS2   VA2NX2N   VA2NX8Y   VA2NX97   VA2NXBV   VA2NXCA   VA2NXCU   VA2NXDP   VA2NXHQ   VA2NXJ6   VA2NXK3   VA2NXKQ   VA2NXKZ   VA2NXLA   VA2P2MD   VA2P2ML   VA2P2MU   VA2P2MZ   VA2P2N4   VA2P2NB   VA2P39D   VA2P39V   VA2P39Y   VA2P3A8   VA2P3AA   VA2P3AK   VA2P4J8   VA2P4JF   VA2P4JS   VA2P4K8   VA2P4MJ   VA2P4WF   VA2P4WV   VA2P4X5   VA2P4ZG   VA2P503   VA2P50E   VA2P50W   VA2P9W3   VA2P9YX   VA2P9ZG   VA2P9ZZ   VA2PA13   VA2PA1Q   VA2PA8W   VA2PA9V   VA2PAA8   VA2PAAP   VA2PABJ   VA2PADJ   VA2PAE8   VA2PCL7   VA2PCM8   VA2PCNW   VA2PCP0   VA2PCP6   VA2PCPS   VA2PCQE   VA2PD2E   VA2PD57   VA2PD6T   VA2PD7J   VA2PD7R   VA2PFK1   VA2PFLU   VA2PFLZ   VA2PG39   VA2PG3N   VA2PG8M   VA2PG8W   VA2PGA4   VA2PGC5   VA2PGCE   VA2PGGZ   VA2PGH0   VA2PGH2   VA2PGJ7   VA2PGP6   VA2PGPW   VA2PGS3   VA2PGSK   VA2PGUT   VA2PGV7   VA2PPDU   VA2PPEN   VA2PPEU   VA2PPFA   VA2PPFB   VA2PPFH   VA2PQ01   VA2PQ0R   VA2PQ18   VA2PQ43   VA2PQ5C   VA2PQ5U   VA2PTY2   VA2PTY5   VA2PTY9   VA2PTYM   VA2PTYZ   VA2PTZ3   VA2PUBM   VA2PUBT   VA2Q575   VA2Q57K   VA2Q57V   VA2Q58C   VA2QEVH   VA2QEVM   VA2QEZ4   VA2QEZA   VA2QFGA   VA2QFHF   VA2QJGC   VA2QJGL   VA2QJGQ   VA2QJGU   VA2QJRL   VA2QJRS   VA2QJW5   VA2QJWB   VA2QWB5   VA2QWB9   VA2QWDY   VA2QWH2   VA2QWHM   VA2QWJ0   VA2QWJK   VA2QWJY   VA2QWLT   VA2QWMM   VA2QWN2   VA2QWN6   VA2QWNG   VA2QWNJ   VA2QWWT   VA2QWX9   VA2QWXF   VA2QWXX   VA2QWY8   VA2QWYV   VA2QX54   VA2QX5H   VA2QX5N   VA2QXAN   VA2QXBP   VA2QXC7   VA2RABG   VA2RABL   VA2RABQ   VA2RABW   VA2RACE   VA2RACH   VA2RAJL   VA2RAJM   VA2RAJN   VA2RAJP   VA2RAJQ   VA2RAJR   VA2RDMC   VA2RDMK   VA2RDN2   VA2RDPY   VA2RDQ8   VA2RDR1   VA2RDRW   VA2RDRZ   VA2RDS2   VA2RDS4   VA2RDS5   VA2RDS6   VA2RDS7   VA2RDS8   VA2RE4D   VA2RE4E   VA2RE4R   VA2RE4T   VA2RE4X   VA2RE57   VA2RE5F   VA2RNM7   VA2RNMF   VA2RNNB   VA2RNNF   VA2RNNG   VA2RNNK   VA2RNNQ   VA2RP22   VA2RP3D   VA2RP5A   VA2RP5Y   VA2RZ12   VA2RZ1K   VA2RZ2G   VA2RZ2T   VA2RZ36   VA2RZ3G   VA2RZ3P   VA2RZ6S   VA2S015   VA2S0KR   VA2S1BU   VA2S1CG   VA2S1D1   VA2S1KD   VA2S1KF   VA2S1KY   VA2S1L5   VA2S1SY   VA2S1T3   VA2S1TZ   VA2S1V2   VA2S1V6   VA2S1VA   VA2S4ZL   VA2S501   VA2S515   VA2S51H   VA2S51L   VA2S51N   VA2S52L   VA2S52N   VA2S52S   VA2S52Y   VA2S53G   VA2S6RH   VA2S6RX   VA2S6SJ   VA2S6U3   VA2S6UB   VA2S70Y   VA2S720   VA2S8UH   VA2S8UY   VA2S8VJ   VA2S8W8   VA2S8WB   VA2S8ZS   VA2S8ZW   VA2S90K   VA2S91X   VA2S92R   VA2S93G   VA2S94M   VA2S9V8   VA2SCTH   VA2SCTN   VA2SCTR   VA2SCU1   VA2SCV3   VA2SDKL   VA2SDKP   VA2SDLB   VA2SDMY   VA2SDP3   VA2SX34   VA2SX4F   VA2SX67   VA2SX6J   VA2SXBN   VA2SXCU   VA2SYRT   VA2SYS4   VA2SYSE   VA2SYSS   VA2SYTB   VA2SYVA   VA2SYYA   VA2SYYB   VA2SYYE   VA2SYYJ   VA2T019   VA2T01G   VA2T03K   VA2T04C   VA2T0C2   VA2TNME   VA2TNMG   VA2TNN7   VA2TNNK   VA2TNNZ   VA2TNP2   VA2TNXK   VA2TNXW   VA2TP2X   VA2TP34   VA2TP4F   VA2TP51   VA2TP9U   VA2TP9X   VA2TPC7   VA2TPD6   VA2TPDQ   VA2TQGN   VA2TQH1   VA2TQKB   VA2TW88   VA2TW95   VA2TW9E   VA2TWNB   VA2TWPC   VA2TWR5   VA2TWRM   VA2TWRX   VA2TWTB   VA2TXZC   VA2TY0Q   VA2TY1A   VA2TY35   VA2TZ0X   VA2TZ19   VA2TZ30   VA2TZ3E   VA2TZ5E   VA2TZ5M   VA2U6R1   VA2U6X1   VA2U6XA   VA2U6XH   VA2U6XP   VA2U6XY   VA2U72Q   VA2U75T   VA2U75W   VA2U75Z   VA2U76B   VA2U76P   VA2UBPS   VA2UBQB   VA2UBQF   VA2UBQH   VA2UBQM   VA2UBR2   VA2UBS5   VA2UBSN   VA2UBTF   VA2UBU9   VA2UBUU   VA2UDXS   VA2UDZM   VA2UE1A   VA2UE61   VA2UE8A   VA2UGPV   VA2UGRH   VA2UGUM   VA2UGVD   VA2UGW7   VA2UGXQ   VA2UHF9   VA2UHFH   VA2UHJ4   VA2UHJF   VA2UK9S   VA2UKA0   VA2UKAD   VA2UKAQ   VA2UKBH   VA2UKBX   VA2ULXB   VA2UM04   VA2UM14   VA2UM1W   VA2UM1Z   VA2UM2B   VA2UM7B   VA2UM8E   VA2UM9D   VA2UM9L   VA2UMC3   VA2UW1Q   VA2UW6A   VA2UW6Z   VA2UX32   VA2UX34   VA2UX3H   VA2UX43   VA2UX4W   VA2UX5G   VA2UX5S   VA2UX5W   VA2UYKL   VA2UYKP   VA2V1PC   VA2V1R7   VA2V1RH   VA2V1RL   VA2V1S1   VA2V1S5   VA2V1S8   VA2V1T5   VA2V1U5   VA2V1UH   VA2V1UM   VA2V9SP   VA2V9U1   VA2V9V3   VA2V9W5   VA2VA59   VA2VA5A   VA2VA5B   VA2VBND   VA2VBP1   VA2VBPG   VA2VBPQ   VA2VBQY   VA2VBW9   VA2VBWD   VA2VBWF   VA2VBWW   VA2VD3Q   VA2VD46   VA2VDBA   VA2VE15   VA2VFFP   VA2VFHT   VA2VFJ1   VA2VFJG   VA2VJLK   VA2VKFP   VA2VKJT   VA2VKKP   VA2VL7Z   VA2VL80   VA2VL83   VA2VL86   VA2VL89   VA2VL8B   VA2VNEP   VA2VNEY   VA2VNEZ   VA2VNF5   VA2VNF7   VA2VNFH   VA2VRNG   VA2VRNZ   VA2VRPE   VA2VRPT   VA2VSW4   VA2VUX0   VA2VUY7   VA2VV0H   VA2VV0K   VA2VV10   VA2VV2X   VA2VWDC   VA2VWDG   VA2VWER   VA2VWEV   VA2VWF2   VA2VWF8   VA2VWWE   VA2VWWU   VA2VZUS   VA2VZV6   VA2VZVK   VA2W02V   VA2W02X   VA2W033   VA2W11G   VA2W12E   VA2W1CC   VA2W1CR   VA2W43H   VA2W443   VA2W44B   VA2W455   VA2W459   VA2W45N   VA2W6VE   VA2WB8K   VA2WB9N   VA2WG8E   VA2WG8H   VA2WG9J   VA2WG9P   VA2WG9R   VA2WG9T   VA2WPL0   VA2WPLH   VA2WPLK   VA2WPLM   VA2WPLW   VA2WPM5   VA2WW4S   VA2WW5M   VA2WW60   VA2WW7F   VA2WW7R   VA2WW85   VA2WW8J   VA2WW8V   VA2WW9C   VA2WWAQ   VA2WWB2   VA2WWB5   VA2WWBF   VA2WWBZ   VA2WXRT   VA2WXS3   VA2X5WM   VA2X60Z   VA2X61K   VA2X62C   VA2X6LH   VA2X6R9   VA2X6RX   VA2X6TG   VA2X6U2   VA2X6UH   VA2X6UM   VA2X6UP   VA2XE76   VA2XE9B   VA2XEAS   VA2XEBU   VA2XEEP   VA2XEF1   VA2XFSF   VA2XFSH   VA2XFSJ   VA2XFSL   VA2XFSV   VA2XK0C   VA2XK0Y   VA2XK16   VA2XK1C   VA2XK1S   VA2XK3W   VA2XKX5   VA2XLSY   VA2XLTJ   VA2XLTU   VA2XLUE   VA2XLVU   VA2XLW8   VA2XLX7   VA2XLXJ   VA2Y0BT   VA2Y0BZ   VA2Y0C7   VA2Y0C9   VA2Y0CA   VA2Y0CE   VA2Y0CJ   VA2Y0CM   VA2Y1DV   VA2Y1E0   VA2Y1EY   VA2Y1F9   VA2Y1H9   VA2Y1HZ   VA2Y1J3   VA2Y1JA   VA2Y1JL   VA2Y1JN   VA2Y1SY   VA2Y1WH   VA2Y5KP   VA2Y5KT   VA2Y5LD   VA2Y5LH   VA2Y5LN   VA2Y5N2   VA2Y5Q7   VA2Y5QM   VA2YAMK   VA2YAP1   VA2YAXC   VA2YB18   VA2YB3B   VA2YB3W   VA2YB3X   VA2YB4C   VA2YBEJ   VA2YBEN   VA2YBEU   VA2YBEW   VA2YBEZ   VA2YBF1   VA2YBF2   VA2YBF3   VA2YKDK   VA2YKE7   VA2YKEQ   VA2YKF3   VA2YKF9   VA2YKFR   VA2YKG3   VA2YKGE   VA2YKJB   VA2YKM9   VA2YLDT   VA2YLGN   VA2YMTF   VA2YMTJ   VA2YN6E   VA2YN6Y   VA2YPJS   VA2YPMD   VA2YPSP   VA2YPSR   VA2YRTW   VA2YRU3   VA2YRUC   VA2YRUQ   VA2YRUV   VA2YRUW   VA2YSUF   VA2YSUR   VA2YSVR   VA2YSVZ   VA2YSW3   VA2YVK8   VA2YVM2   VA2YVND   VA2YVQ5   VA2YVQX   VA2YVR1   VA2YVR9   VA2YVRF   VA2YVRG   VA2YVRK   VA2YVS5   VA2YXJ2   VA2YXJA   VA2YXJD   VA2YXRP   VA2YXVH   VA2YYZB   VA2YZ02   VA2YZ4R   VA2YZ4S   VA2YZ56   VA2YZ5B   VA2YZ5J   VA2YZ5X   VA2Z4UT   VA2Z4V1   VA2Z4V9   VA2Z4VB   VA2Z4VC   VA2Z4VN   VA2Z8GS   VA2Z8JH   VA2Z8MH   VA2Z8PD   VA2Z8RS   VA2ZA6C   VA2ZA71   VA2ZA9C   VA2ZA9H   VA2ZA9N   VA2ZAA5   VA2ZAAG   VA2ZL9C   VA2ZL9G   VA2ZL9H   VA2ZL9U   VA2ZLA0   VA2ZLDV   VA2ZLG0   VA2ZLG9   VA2ZMM5   VA2ZMR1   VA2ZN4Z   VA2ZN50   VA2ZN51   VA2ZN52   VA2ZT9J   VA2ZT9P   VA2ZT9S   VA2ZT9X   VA2ZT9Y   VA2ZY12   VA2ZY13   VA2ZY9S   VA2ZYB5   VA2ZYBD   VA2ZYBV   VA2ZYFJ   VA2ZYFM   VA2ZYFQ   VA2ZYFS   VA3017J   VA3017M   VA3017N   VA3017Q   VA30195   VA30196   VA308JF   VA308K3   VA308K7   VA308KK   VA309BH   VA309BS   VA309BX   VA309C7   VA30BG3   VA30BLW   VA30CH5   VA30CHZ   VA30CKQ   VA30CP8   VA30CQ6   VA30CQB   VA30DD4   VA30DDB   VA30E3Q   VA30E3Y   VA30ERW   VA30ESW   VA30ETN   VA30EUC   VA30EXH   VA30EYP   VA30LQR   VA30LRU   VA30LS3   VA30LS6   VA30LSJ   VA30LT7   VA30MS5   VA30MS7   VA30MSH   VA30MUA   VA30P8H   VA30PAN   VA30PDY   VA30PEZ   VA30PL2   VA30PNL   VA30PP3   VA30WRQ   VA30WRR   VA30WRS   VA30WRT   VA30WRU   VA30WRV   VA30ZMP   VA30ZMU   VA30ZNG   VA30ZNX   VA30ZP6   VA30ZPX   VA30ZSN   VA30ZT4   VA311DM   VA311DX   VA3185T   VA3185U   VA3189M   VA3189N   VA318KK   VA318KZ   VA318L9   VA318LL   VA318LT   VA318M1   VA31JVE   VA31JVK   VA31K9P   VA31K9Q   VA31K9R   VA31K9S   VA31K9T   VA31K9U   VA31KG9   VA31KGA   VA31KGB   VA31LJZ   VA31LK5   VA31LK7   VA31LN5   VA31LNK   VA31LNQ   VA31LNS   VA31LNU   VA31LNV   VA31LP1   VA31LP4   VA31LP9   VA31MZ9   VA31MZB   VA31MZC   VA31MZE   VA31MZG   VA31NLS   VA31SVZ   VA31SW0   VA31SW3   VA31SWZ   VA31TFR   VA31THX   VA31VE9   VA31VEA   VA31VEB   VA31VEC   VA31VED   VA31VUC   VA31VZ9   VA31WX7   VA31WXM   VA321PS   VA321PT   VA321PU   VA321QA   VA321QB   VA3226N   VA3226P   VA322AL   VA322AM   VA322DC   VA322DD   VA326VL   VA326VS   VA326W8   VA326WB   VA326WQ   VA326WT   VA326X7   VA326XB   VA326XE   VA326XZ   VA327G5   VA328BZ   VA328C0   VA328Z1   VA328Z8   VA32B1R   VA32B1U   VA32FZG   VA32FZH   VA32FZJ   VA32FZL   VA32FZM   VA32FZP   VA32FZW   VA32G0Q   VA32G0T   VA32G15   VA32GXM   VA32H8N   VA32H8P   VA32H94   VA32H96   VA32LKD   VA32LPQ   VA32LQB   VA32LQU   VA32LRR   VA32LRW   VA32NFC   VA32NFD   VA32NFE   VA32NFF   VA32NHJ   VA32NHK   VA32Q6J   VA32Q6K   VA32QGR   VA32QGV   VA32QGX   VA32QH2   VA32T05   VA32T09   VA32T3V   VA32T41   VA32T4G   VA331AR   VA331BF   VA331C1   VA331C3   VA331DW   VA331EG   VA331Y0   VA332XV   VA334AP   VA334AQ   VA334P8   VA334PF   VA33508   VA33509   VA336RZ   VA337BC   VA337BE   VA337CV   VA337CW   VA3399N   VA33BMK   VA33BTS   VA33BUA   VA33BVX   VA33BWN   VA33BY2   VA33C1W   VA33C21   VA33DH5   VA33DHN   VA33DTG   VA33DTM   VA33FKU   VA33FLC   VA33FLG   VA33K80   VA33L1L   VA33L1M   VA33L1Y   VA33L1Z   VA33L4F   VA33L4L   VA33NUA   VA33NUP   VA33NUZ   VA33Y3N   VA33Y49   VA33Y4P   VA33Y5L   VA33Y5N   VA33Y5S   VA344NZ   VA344P2   VA344P5   VA344PJ   VA344PN   VA344Q1   VA345KX   VA345KZ   VA3477Q   VA34789   VA347AZ   VA347C5   VA348X8   VA348Y0   VA34D3A   VA34D3N   VA34D3V   VA34D3X   VA34D3Z   VA34D42   VA34GR2   VA34GRV   VA34GS8   VA34GSJ   VA34GSK   VA34GSP   VA34K4L   VA34K4S   VA34K61   VA34K6D   VA34NA5   VA34NA8   VA34NBW   VA34NBX   VA34NC0   VA34NC4   VA34P48   VA34P5S   VA34P68   VA34P76   VA34P79   VA34P7C   VA34QNU   VA34QP7   VA34QXR   VA34QXY   VA34SHX   VA34SHZ   VA34SJ4   VA34SJ5   VA34U54   VA34U5W   VA34U6M   VA34U9H   VA34UBK   VA34UXR   VA34XYF   VA34XYM   VA34XZJ   VA34Y1C   VA34Y1W   VA350MC   VA350MH   VA350N6   VA350N9   VA350NB   VA350ND   VA352YK   VA352ZN   VA3530X   VA35311   VA35314   VA3531D   VA354QE   VA354QV   VA354RE   VA354RU   VA35503   VA35509   VA356JP   VA356M3   VA356MR   VA356UJ   VA356VE   VA3579A   VA359BS   VA359C6   VA359NX   VA359NZ   VA359P1   VA359PB   VA35ASM   VA35ATX   VA35AWC   VA35AYF   VA35EQT   VA35EQW   VA35ERB   VA35ES0   VA35JLR   VA35JLY   VA35JML   VA35JND   VA35JNH   VA35JNK   VA35MFG   VA35MFH   VA35MFJ   VA35MFK   VA35MFN   VA35NXF   VA35NZ1   VA35NZZ   VA35P0D   VA35P0H   VA35P0V   VA35P0X   VA35S6A   VA35S6H   VA35S6L   VA35S6Q   VA35S6R   VA35TUR   VA35TUS   VA35TUT   VA35TUV   VA35V8B   VA35V8D   VA35V8Y   VA35V90   VA35V97   VA35V9C   VA35VY2   VA35VY8   VA35VYH   VA35VYP   VA35VYQ   VA35VYV   VA35Y35   VA35Y36   VA35Y3S   VA35Y3T   VA360DV   VA360DW   VA360E0   VA360EA   VA360ED   VA360FP   VA362PM   VA362PN   VA362RN   VA362RP   VA362UQ   VA362V1   VA366MH   VA366ML   VA366S3   VA366S5   VA366SN   VA366TC   VA368H6   VA368H9   VA36C9C   VA36C9D   VA36C9V   VA36C9W   VA36CA0   VA36CA1   VA36FD0   VA36FD3   VA36FDA   VA36FDE   VA36FDK   VA36FE1   VA36JDW   VA36JE4   VA36JMT   VA36JNA   VA36JND   VA36JNQ   VA36K3J   VA36K3K   VA36K9D   VA36KGU   VA36KH0   VA36KH1   VA36N6M   VA36NEB   VA36NEC   VA36NQN   VA36QMX   VA36QMY   VA36QPG   VA36QQ9   VA36QQZ   VA36QRV   VA36QRX   VA36V2Y   VA36V5S   VA36V5T   VA36V5U   VA36XQG   VA36XQK   VA36XR8   VA36XRF   VA36XRT   VA36XS0   VA36Z5A   VA36Z5D   VA36Z5H   VA36ZBC   VA36ZBD   VA36ZBR   VA3705Q   VA3705X   VA37066   VA3706W   VA3707Y   VA3708B   VA376SR   VA376SS   VA376T5   VA376T6   VA376TA   VA376TC   VA37A02   VA37A07   VA37A0S   VA37A18   VA37F22   VA37F26   VA37F27   VA37F28   VA37F29   VA37F2A   VA37KD3   VA37KGK   VA37KGN   VA37KGU   VA37KGY   VA37KHA   VA37NUS   VA37NUU   VA37NUX   VA37NVS   VA37NVT   VA37NW4   VA37NWA   VA37P3D   VA37P3E   VA37P5G   VA37P5H   VA37PAL   VA37PAT   VA37TVC   VA37TVE   VA37TVW   VA37TW7   VA37TWA   VA37TWB   VA37TWC   VA37U3K   VA37U3Q   VA37U3R   VA37U3S   VA37UAS   VA37UAT   VA37Z7R   VA37Z8Q   VA37Z8T   VA37Z90   VA37ZAH   VA37ZAQ   VA383AS   VA383BP   VA383BY   VA383BZ   VA3847B   VA38489   VA3848S   VA3848T   VA3848U   VA3848V   VA387YX   VA387Z0   VA387Z4   VA387Z9   VA38DJE   VA38DJG   VA38DJJ   VA38DJN   VA38DJQ   VA38DJR   VA38G2M   VA38G32   VA38G3F   VA38G3M   VA38GHW   VA38GHX   VA38GJY   VA38GJZ   VA38N46   VA38N4B   VA38N4P   VA38N5Z   VA38PGW   VA38PGY   VA38UDS   VA38UDV   VA38UDX   VA38UDZ   VA38UE1   VA38Y7V   VA38Y7W   VA38YD7   VA38YKM   VA38YL2   VA38YL3   VA3913K   VA3913M   VA3913P   VA39141   VA394UN   VA394VH   VA394W1   VA3956H   VA3956N   VA3956U   VA3963Q   VA3963S   VA3963Y   VA39646   VA39AG5   VA39AHY   VA39AJ0   VA39B7T   VA39B80   VA39B88   VA39B8J   VA39B8P   VA39B8X   VA39G3N   VA39G3P   VA39G3R   VA39G5D   VA39G5H   VA39G5J   VA39GXQ   VA39GZE   VA39H00   VA39H02   VA39L2K   VA39L2L   VA39L2X   VA39L31   VA39L38   VA39L3B   VA39MJA   VA39MJK   VA39MJM   VA39MK3   VA39MK6   VA39P98   VA39P9D   VA39SLZ   VA39SM4   VA39SMG   VA39SMQ   VA39WQ4   VA39WQ5   VA39X01   VA39X02   VA39X0R   VA39X0T   VA39YMF   VA39YMH   VA39YTM   VA39YTN   VA39YTP   VA39YTQ   VA3A060   VA3A061   VA3A062   VA3A063   VA3A064   VA3A065   VA3A08C   VA3A08D   VA3A08E   VA3A08F   VA3A08Z   VA3A09A   VA3A227   VA3A22F   VA3A22K   VA3A22Q   VA3A22U   VA3A3MR   VA3A3MU   VA3A3MY   VA3A3MZ   VA3A3N0   VA3A3N1   VA3A428   VA3A439   VA3A495   VA3A497   VA3A49H   VA3A49L   VA3A4A0   VA3A4A6   VA3A503   VA3A509   VA3A50G   VA3A50K   VA3A50N   VA3A50Q   VA3A5E8   VA3A5J2   VA3A5NX   VA3A5NZ   VA3A5P1   VA3A5P3   VA3A5P7   VA3A5QT   VA3A69Z   VA3A6A1   VA3A6A9   VA3A6AA   VA3A6AE   VA3A6AJ   VA3A7WW   VA3A817   VA3A81M   VA3A825   VA3A83E   VA3A84G   VA3A9RP   VA3A9RU   VA3A9RZ   VA3A9SQ   VA3A9SV   VA3A9SZ   VA3ACSP   VA3ACSV   VA3ACT5   VA3ACZ3   VA3AD0J   VA3AD39   VA3AEK6   VA3AEKP   VA3AEL5   VA3AENB   VA3AENG   VA3AENT   VA3AGRH   VA3AGRK   VA3AGT1   VA3AGY5   VA3AGYF   VA3AGZC   VA3AJ6B   VA3AJ6L   VA3AJ96   VA3AJ97   VA3ALEY   VA3ALEZ   VA3ALMH   VA3ALMJ   VA3ALNS   VA3ALNU   VA3ANFL   VA3ANKD   VA3ANKZ   VA3AQ01   VA3AQ0B   VA3AQ0C   VA3ARMG   VA3ARMM   VA3ARMW   VA3AT4L   VA3AT56   VA3AUJA   VA3AUJB   VA3AYPR   VA3AYQU   VA3AYRD   VA3B09A   VA3B3DC   VA3B4ZL   VA3B5UX   VA3B6G6   VA3B6G9   VA3B93P   VA3BBAA   VA3BCWB   VA3BCXJ   VA3BG5R   VA3BKX1   VA3BL5P   UDI-DI 00763000587062    Lots VA2JLZT   VA2JMYG   VA2JQGQ   VA2NCB1   VA2NN7R   VA2NN95   VA2NN9V   VA2P2N4   VA2P9VM   VA2PA8W   VA2PCP0   VA2S1KD   VA2S1KF   VA2S1V2   VA2SCUY   VA30ZT4   VA31KGB   VA31LNS   VA31LNV   VA31LP1   VA31LP4   VA31MZB   VA31SVZ   VA31VE9   VA321PU   VA322AM   VA322DD   VA34D3Z   VA34K6D   VA34P48   VA34U6M   VA34U9H   VA350N9   VA35509   VA356M3   VA35JLY   VA35JML   VA36Z5H   VA376T5   VA37KGK   VA37P5G   VA37TVW   VA383BP   VA3848V   VA38GHX   VA38N4P   VA38YL3   VA3A22U   VA3A3MU;    3560030    UDI-DI 00763000203887    Lots VA24HE0   VA24Q0F   VA24QED   VA24QEF   VA24WJD   VA25181   VA25182   VA26CMD   VA26CMZ   VA26CN0   VA26CN1   VA26CN3   VA26E7A   VA26EGF   VA26EGG   VA275QF   VA275QG   VA2760G   VA276AW   VA276AX   VA276GM   VA276GN   VA276GP   VA277MV   VA277PR   VA277V8   VA277V9   VA277VU   VA277VV   VA277VW   VA27CCU   VA27CGB   VA27CGC   VA27CGD   VA27D43   VA27D44   VA27D5W   VA27UGH   VA27UW3   VA27YHA   VA27YKP   VA280D2   VA280MV   VA28488   VA2871D   VA2872X   VA28BHF   VA28BHG   VA28BLL   VA28BQ1   VA28BR1   VA28BR2   VA28K4U   VA28Z67   VA28Z6B   VA28Z6N   VA28Z6P   VA28Z6X   VA28Z9M   VA29ES2   VA29ES3   VA29ES4   VA29FCW   VA29MH7   VA29MKB   VA29MTM   VA29N1U   VA29ZUL   VA2A2Z8   VA2A2Z9   VA2A2ZA   VA2AEYH   VA2ALWG   VA2APTS   VA2ARCX   VA2B686   VA2B687   VA2B6NC   VA2B8H1   VA2B8HP   VA2BCYM   VA2BD3Z   VA2BG7Z   VA2BG83   VA2BG87   VA2BG8X   VA2BG9P   VA2BLFF   VA2BLFH   VA2BLFJ   VA2BLFK   VA2BM7Q   VA2BT60   VA2BU2P   VA2BUMF   VA2C5LL   VA2C7N9   VA2CC3M   VA2CCF6   VA2CCJT   VA2CSQV   VA2CSQW   VA2CSQX   VA2CSQY   VA2CT5C   VA2CUNA   VA2CZT5   VA2CZT6   VA2D8TH   VA2DDM4   VA2DV4U   VA2DV70   VA2DV94   VA2E1Q6   VA2E1QA   VA2E4AW   VA2E4D1   VA2EGXS   VA2EGXT   VA2EGZB   VA2EJF6   VA2EMNG   VA2EMPB   VA2EP63   VA2EP76   VA2F0A5   VA2F0MJ   VA2F0PS   VA2F0R7   VA2F0S7   VA2F32V   VA2F349   VA2F34L   VA2F37J   VA2FBWW   VA2FBXW   VA2FBYC   VA2FBYK   VA2FBYS   VA2FJ04   VA2FJ0E   VA2FJ3T   VA2FJDG   VA2FJDK   VA2FJDP   VA2FJU0   VA2FL2Z   VA2FL90   VA2FL99   VA2FLA7   VA2FLA9   VA2FLAG   VA2FLPL   VA2FLSF   VA2FLSN   VA2FLTG   VA2FLV1   VA2FLVY   VA2FM0B   VA2FMQY   VA2FMRW   VA2FMSW   VA2FMTT   VA2FP7E   VA2FP7G   VA2FP83   VA2FPAV   VA2GKGD   VA2GT10   VA2HET9   VA2HEUH   VA2HEUM   VA2HEUW   VA2HT55   VA2HT5F   VA2J72L   VA2J8ZF   VA2JLWF   VA2JLX7   VA2JLXB   VA2JLXE   VA2JTN9   VA2KVRP   VA2L1XV   VA2L1Y3   VA2L1Y7   VA2L1YD   VA2L7LK   VA2L7M1   VA2LGZK   VA2LH09   VA2LH1X   VA2LH1Z   VA2LWMF   VA2LWN1   VA2M0HN   VA2M0K2   VA2M7EH   VA2M7EP   VA2M7F4   VA2M7F8   VA2PD3B   VA2PD41   VA2PPFK   VA2PPGP   VA2QESR   VA2QET4   VA2QETF   VA2QEUE   VA2QYGH   VA2QYJT   VA2QYK0   VA2RAD2   VA2RP80   VA2RPBR   VA2T1BR   VA2T1HA   VA2TSXS   VA2TSYM   VA2TSYR   VA2TSZ3   VA2TSZB   VA2TSZF   VA2UW75   VA2UW84   VA2VKHF   VA2VKJM   VA2W6JR   VA2WPMQ   VA2WPMS   VA2YBF4   VA2YVJA   VA2Z4UG   VA2ZMK3   VA2ZMKC   VA30BN3   VA30BNV   VA30PG1   VA30PGX   VA31985   VA32NHL   VA32NHN   VA331Y9   VA332Y2   VA332Y7   VA332YQ   VA334P0   VA34Y1Z   VA35MC0   VA35Y60   VA35Y63   VA37A1N   VA3839X   VA383A3   VA38UE5   VA3913T   VA3ARMA  UDI-DI 00763000587000    Lots VA2F0R7   VA2GT10   VA2LH1Z   VA2M7F8   VA2QET4   VA2QEUE   VA2T1HA   VA2TSXS   VA2TSYM   VA2VKJM   VA2WPMQ   VA2WPMS   VA2YVJA   VA2Z4UG   VA334P0",
      "more_code_info": ""
    }
  ]
}