  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Springfield",
      "state": "NJ",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95429",
      "recalling_firm": "DRG International, Inc.",
      "address_1": "841 Mountain Ave",
      "address_2": "N/A",
      "postal_code": "07081-3437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "AZ, CA,  NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam",
      "recall_number": "Z-0258-2025",
      "product_description": "17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in serum or plasma   Model/Catalog Number: EIA-1292",
      "product_quantity": "504 units",
      "reason_for_recall": "The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.",
      "recall_initiation_date": "20240913",
      "center_classification_date": "20241029",
      "report_date": "20241106",
      "code_info": "Lot Code: 04048474012927 (DRG Instruments GmbH) lots: 22K024 and 22K034",
      "more_code_info": ""
    }
  ]
}