{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "800 Macarthur Blvd",
      "reason_for_recall": "NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature  2. NIBP hose blockage identification function",
      "address_2": "",
      "product_quantity": "14 units",
      "code_info": "NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5",
      "center_classification_date": "20191029",
      "distribution_pattern": "US Nationwide distributions.",
      "state": "NJ",
      "product_description": "Rosie4 Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP). Temperature (TEMP) and Carbon Dioxide (CO2) on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians",
      "report_date": "20191106",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Mindray DS USA, Inc. dba Mindray North America",
      "recall_number": "Z-0237-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83979",
      "termination_date": "20210806",
      "more_code_info": "",
      "recall_initiation_date": "20191001",
      "postal_code": "07430-2001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}