  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97479",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.  US Nationwide including Guam and Puerto Rico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Burundi, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guatemala, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Mali, Malta, Mauritania, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Sierra Leone, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Timor-Leste, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zambia, and Zimbabwe.",
      "recall_number": "Z-0236-2026",
      "product_description": "BD BACTEC Blood Culture System, labeled as the following:    1. BD BACTEC FX, Instrument Top, Packaged, Catalog Number: 44138509.    2. Instrument Top Remanufactured Bactec Fx, Catalog Number: 441676.     3. BACTEC 9240 System, Catalog Number: 445569.    4. BD BACTEC FX40 Instrument, Catalog Number: 44229609.",
      "product_quantity": "177 units",
      "reason_for_recall": "Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.",
      "recall_initiation_date": "20250923",
      "center_classification_date": "20251023",
      "report_date": "20251029",
      "code_info": "1. Catalog Number: 44138509; UDI/DI: 00382904413859; Serial Numbers: FT5675, FT5497, FT4948, FT5781, FT5120, FT3993, FT7779, FT3387, FT1993, FT5019, FT9109, FT8412, FT7176, FT7344, FT4282, FT7400, FT6093, FT0962, FT2780, FT0795, FT0794, FT2799, FT5924, FT5279, FT5286, FT0572, FT4931, FT7289, FT9801, FT2652, FT5232, FT1348, FT8083, FT0640, FT7159, FT7034, FT8410, FT0098, FT2650, FT6203, FT3480, FT4917, FT3031, FT3033, FT4902, FT7189.  2. Catalog Number: 441676; UDI/DI: 00382904416768; Serial Numbers: FT7439A, FT7201A, FT5679, FT0599, FT0598, FT0644, FT0600, FT5025, FT2845, FT0525, FT0715, FT4439, FT0330, FT2313, FT3020, FT3022, FT0745, FT5090, FT4340, FT6225, FT4977, FT0571, FT0570, FT0498, FT5248, FT5270, FT3537, FT4448, FT3538, FT3486, FT0340, FT1517, FT5590, FT3030, FT4800, FT1774, FT2263, FT5014, FT5013, FT7184A.   3. Catalog Number: 445569; UDI/DI: 00382904455699; Serial Numbers: DB4372, DB6043, DB5030, DB5825, DB2835, DB3251, DB3245, DB3145, DB5876, DB5906, DB3143, DB3204, DB6891, PM4385R, DB4718, DB3231, DB2786, DB6628, DB6393, DB6936, DB4987, DB5347A, DB6123, DB6281, DB6933, DB6451, DB5602, DB5858, DB4763, DB6302, DB6165, DB5596, DB5089, DB7157, DB5021, DB3086, DB4776, DB6296, DB5328, DB2245, DB2483, DB5329, DB2946, DB2945, DB4348, DB5319, DB4299, DB4484, MD2033, DB3062, RC1164, DB6323, DB4550, DB6213, DB5083, DB6383, DB6381, DB6457, DB6169, DB6212, DB6929, DB6978, DB6913, DB6540, DB6699, DB6989.  4. Catalog Number: 44229609; UDI/DI: 00382904422967; Serial Numbers: FF1979, FF0713, FF0410, FF1561, FF2832, FF4395, FF1500, FF0697, FF0933, FF1499, FF1025, FF5511, FF0027, FF1118, FF3195, FF0115, FF5979, FF6535, FF0391, FF3284, FF0772, FF0570, FF0638, FF1599, FF1271.",
      "more_code_info": ""
    }
  ]
}