  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Maryland Heights",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78665",
      "recalling_firm": "Elekta Inc",
      "address_1": "13723 Riverport Dr Ste 100",
      "address_2": "N/A",
      "postal_code": "63043-4819",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of:  VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of:    AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PERU, PHILIPPINES, PORTUGAL, RUSSIA,   SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, TAIWAN, THAILAND, TURKEY, VENEZUELA, and VIETNAM",
      "recall_number": "Z-0234-2018",
      "product_description": "Elekta ERGO++    Product Usage:  ERGO is often used for stereotactic treatments.",
      "product_quantity": "182",
      "reason_for_recall": "Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.",
      "recall_initiation_date": "20170901",
      "center_classification_date": "20171212",
      "termination_date": "20191105",
      "report_date": "20171220",
      "code_info": "Version 1.7.3 and higher"
    }
  ]
}