  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93052",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution",
      "recall_number": "Z-0194-2024",
      "product_description": "EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190",
      "product_quantity": "3948 devices",
      "reason_for_recall": "There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.",
      "recall_initiation_date": "20231012",
      "center_classification_date": "20231109",
      "report_date": "20231115",
      "code_info": "Model BF-P190 UDI-DI: 4953170342110",
      "more_code_info": ""
    }
  ]
}