{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95438",
      "recalling_firm": "Smiths Medical ASD, Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-0191-2025",
      "product_description": "medex TranStar 60in (152cm) Single Monitoring Kit 10/EA, Product Code REF MX9504T",
      "product_quantity": "5880 units",
      "reason_for_recall": "Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.",
      "recall_initiation_date": "20240924",
      "center_classification_date": "20241023",
      "report_date": "20241030",
      "code_info": "UDI/DI 50351688504034 (case), 10351688504036 (pouch), Lot Numbers: 4277662, 4287634, 4289405",
      "more_code_info": ""
    }
  ]
}