{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95438",
      "recalling_firm": "Smiths Medical ASD, Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-0182-2025",
      "product_description": "medex TranStar Single OR Set 1/EA, Product Code REF MX20846R2",
      "product_quantity": "260 units",
      "reason_for_recall": "Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.",
      "recall_initiation_date": "20240924",
      "center_classification_date": "20241023",
      "report_date": "20241030",
      "code_info": "UDI/DI 50351688508780 (case), 10351688508782 (pouch) , Lot Numbers:  4282732, 4287721",
      "more_code_info": ""
    }
  ]
}