  "results": [
    {
      "status": "Ongoing",
      "city": "Le Noirmont",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86419",
      "recalling_firm": "Bien-Air Surgery Sa",
      "address_1": "rue de l' ouest 2b",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of CA, MI and MN.",
      "recall_number": "Z-0176-2021",
      "product_description": "Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.",
      "product_quantity": "Total US = 1120 foot pedals",
      "reason_for_recall": "The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.",
      "recall_initiation_date": "20200911",
      "center_classification_date": "20201021",
      "report_date": "20201028",
      "code_info": "Model 1600407 / Catalog number 1600407-001  Serial Numbers:  16C0017  16F0013  16I0009  16K0017  17F0039  17I0015  17L0013"
    }
  ]
}