      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99886",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0175-2027",
      "product_description": "Medline surgical convenience kits labeled as:     1) CV BASIN PACK-LF, Medline Product number/SKU DYNJ0538615K;   2) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615J;   3) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615L.",
      "product_quantity": "235 kits",
      "reason_for_recall": "Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.",
      "recall_initiation_date": "20260902",
      "center_classification_date": "20261015",
      "report_date": "20261021",
      "code_info": "Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 26BBA206;  Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25LBQ023;  Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25KBR149;  Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25EBF588;  Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25ABI296;  Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 24IBP190;  Medline Product Number/SKU DYNJ0538615J; UDI/DI 10195327673376 (each) 40195327673377 (case), Lot number 24DBC864.",
      "more_code_info": ""
    }
  ]
}