  "results": [
    {
      "status": "Ongoing",
      "city": "Grimes",
      "state": "IA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97659",
      "recalling_firm": "American Contract Systems Inc.",
      "address_1": "1601 Se Gateway Dr Ste 120",
      "address_2": "",
      "postal_code": "50111-5223",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of SD, IA, MN, WA, IL.",
      "recall_number": "Z-0170-2026",
      "product_description": "Basic Biopsy Tray, Item Number NMBP44L",
      "product_quantity": "32,433 total",
      "reason_for_recall": "Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.",
      "recall_initiation_date": "20250904",
      "center_classification_date": "20251016",
      "report_date": "20251022",
      "code_info": "UDI-DI 191072188120 Lot 8515411 Entire lot impacted",
      "more_code_info": ""
    }
  ]
}