{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-21",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99883",
      "recalling_firm": "Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.",
      "recall_number": "Z-0166-2027",
      "product_description": "Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.",
      "product_quantity": "359 trays",
      "reason_for_recall": "Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.",
      "recall_initiation_date": "20260916",
      "center_classification_date": "20261015",
      "report_date": "20261021",
      "code_info": "Catalog Number: 6543-6-514. GTIN:  07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.",
      "more_code_info": ""
    }
  ]
}