      "status": "Ongoing",
      "city": "Miami",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99832",
      "recalling_firm": "Beckman Coulter, Inc.",
      "address_1": "11800 Sw 147th Ave",
      "address_2": "",
      "postal_code": "33196-2500",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.",
      "recall_number": "Z-0161-2027",
      "product_description": "IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997;  IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684;  IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683",
      "product_quantity": "",
      "reason_for_recall": "The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement  to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual  preventive maintenance (PM) requirements without adequate reliability evaluation. When the check  valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor,  potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure),  biohazardous exposure, property damage, and delayed patient results.",
      "recall_initiation_date": "20260812",
      "center_classification_date": "20261015",
      "report_date": "20261021",
      "code_info": "Product/UDI-DI:  IQ200 SPRINT/10837461001256, IQ200 SELECT/10837461001300, IQ200 ELITE/10837461001355.  All Serial Numbers",
      "more_code_info": ""
    }
  ]
}