  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-21",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99804",
      "recalling_firm": "Steris Corporation",
      "address_1": "5960 Heisley Rd",
      "address_2": "",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-0156-2027",
      "product_description": "RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117",
      "product_quantity": "106758 units",
      "reason_for_recall": "If the vented cap becomes blocked, gas may accumulate  and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.",
      "recall_initiation_date": "20260921",
      "center_classification_date": "20261014",
      "report_date": "20261021",
      "code_info": "UDI: 00724995245962 Lot numbers with expiration dates: 688019 10/19/2027 688021 10/19/2027 688023 10/19/2027 688025 10/19/2027 688576 10/26/2027 688580 10/26/2027 688582 10/26/2027 688584 10/26/2027 688586 10/26/2027 688588 10/26/2027 689173 11/2/2027 689175 11/2/2027 689177 11/2/2027 689179 11/2/2027 689983 11/10/2027 689985 11/10/2027 689987 11/10/2027 690299 11/16/2027 690301 11/16/2027 690995 11/24/2027 690997 11/24/2027 690999 11/24/2027 691001 11/24/2027 691741 12/1/2027 691743 12/1/2027 691745 12/8/2027 692203 12/8/2027 692206 12/8/2027 692208 12/8/2027 692210 12/8/2027 692213 12/8/2027 692215 12/8/2027 692217 12/9/2027 692572 12/10/2027 692574 12/10/2027 692576 12/10/2027 692578 12/10/2027 692580 12/10/2027 692582 12/10/2027 692730 12/14/2027 692732 12/14/2027 692734 12/14/2027 692736 12/14/2027 692738 12/14/2027 692740 12/14/2027 692742 12/14/2027 692744 12/14/2027 694359 12/30/2027 694361 12/30/2027 694363 12/30/2027",
      "more_code_info": ""
    }
  ]
}