{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Miami",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63302",
      "recalling_firm": "Nipro Medical Corporation",
      "address_1": "3150 Nw 107th Ave",
      "address_2": "N/A",
      "postal_code": "33172",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-Distributed in the state of New York and the countries of Spain, Germany, Argentina, Panama, Mexico, Chile, Colombia, Ecuador, Peru, and Brazil.",
      "recall_number": "Z-0137-2013",
      "product_description": "AquaLiner Hydrophilic Guidewire 0.035\" X 260cm, Straight/ Stiff.    AquaLiner Hydrophilic Guidewire is designed for use in angiographic procedures to introduce/ position catheters and interventional devices within the vasculature.",
      "product_quantity": "300",
      "reason_for_recall": "Nipro Medical Corporation, Miami, FL recalled their AquaLiner Hydrophilic Guidewire because the wires contain insufficient lubricity which causes them to get stuck in the housing and become nonfunctional.",
      "recall_initiation_date": "20120727",
      "center_classification_date": "20121025",
      "termination_date": "20130525",
      "report_date": "20121031",
      "code_info": "Item Code: AL+60000003  Lot #: 11F20, 11I27, 12B10"
    }
  ]
}