  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95380",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "N/A",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.",
      "recall_number": "Z-0135-2025",
      "product_description": "Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue  Model/Catalog Number: TFL-SLS",
      "product_quantity": "159 units",
      "reason_for_recall": "The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI),  a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.",
      "recall_initiation_date": "20240813",
      "center_classification_date": "20241018",
      "report_date": "20241030",
      "code_info": "UDI  00821925044135 All Serial Numbers",
      "more_code_info": ""
    }
  ]
}