  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95380",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "N/A",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.",
      "recall_number": "Z-0134-2025",
      "product_description": "Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue,  Model/Catalog Number: TFL-PLS",
      "product_quantity": "1373 units",
      "reason_for_recall": "The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI),  a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.",
      "recall_initiation_date": "20240813",
      "center_classification_date": "20241018",
      "report_date": "20241030",
      "code_info": "UDI  00821925044111 All Serial Numbers:",
      "more_code_info": ""
    }
  ]
}