  "results": [
    {
      "status": "Terminated",
      "city": "Williamston",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83774",
      "recalling_firm": "Centurion Medical Products Corporation",
      "address_1": "100 Centurion Way",
      "address_2": "N/A",
      "postal_code": "48895-9086",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0121-2020",
      "product_description": "Centurion  CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant.  ReOrder: 330CR",
      "product_quantity": "204 each",
      "reason_for_recall": "A  mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.",
      "recall_initiation_date": "20190828",
      "center_classification_date": "20191010",
      "termination_date": "20200723",
      "report_date": "20191016",
      "code_info": "Batch/Lot:  2019061201"
    }
  ]
}