  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83425",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-0114-2020",
      "product_description": "Aestiva MRI, Model Numbers:  a) 1006-9310-000  b) 1006-9110-000  c) 1006-9023-000  d) 1006-9028-000  e) 1006-9310-000-305077  f) 1006-9310-000-015243  g) 1006-9310-000-017602  h) 1006-9310-000-103785  I) 1006-9310-000-025109  j) 1006-9310-000-009650  k) 1006-9310-000-015224  l) 1006-9310-000-031881  m) 1006-9310-000-031854  n) 1006-9310-000-026571",
      "product_quantity": "502 devices",
      "reason_for_recall": "Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network.  An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.",
      "recall_initiation_date": "20190719",
      "center_classification_date": "20191217",
      "termination_date": "20260413",
      "report_date": "20191225",
      "code_info": "a) 1006-9310-000  Lot/Serial Numbers:    AMTE00190  AMTH00218  AMTE00125  AMTE00126  AMTE00118  AMTG00102  AMTF00224  AMTG00108    b) 1006-9110-000  Lot/Serial Numbers:    AMTG00211  AMTG00149  AMTF00152  AMTH00115  AMTH00172  AMTH00108  AMTF00127  AMTH00125  AMTF00128  AMTF00189  AMTH00100  AMTG00111  AMTH00210  AMTG00178  AMTG00199  AMTF00126  AMTF00153  AMTG00113  AMTH00134  AMTF00168  AMTH00114  AMTG00177  AMTF00211  AMTF00166  AMTF00197  AMTF00143  AMTG00156  AMTG00115  AMTE00221  AMTF00125  AMTG00146  AMTG00167  AMTG00179  AMTG00190  AMTH00193  AMTH00194  AMTF00167  AMTF00141  AMTH00116  AMTG00169  AMTG00159  AMTG00212  AMTG00213  AMTF00144  AMTG00112  AMTH00126  AMTG00166  AMTG00155  AMTF00165  AMTE00222  AMTF00210  AMTE00225  AMTG00197  AMTG00148  AMTH00118  AMTH00133  AMTF00142  AMTF00129  AMTH00190  AMTE00227  AMTG00147  AMTG00201  AMTG00175  AMTF00169  AMTF00212  AMTG00165  AMTF00154  AMTG00145  AMTF00203  AMTH00119  AMTF00151  AMTH00117  AMTG00168  AMTG00176  AMTF00150  AMTF00209  AMTH00186  AMTF00140  AMTF00108  AMTE00223  AMTH00132  AMTG00158  AMTH00107  AMTF00183  AMTG00114  AMTG00200  AMTG00198    c) 1006-9023-000  Lot/Serial Numbers:    AMTE00239  AMTF00120  AMTE00155  AMTH00183  AMTH00144  AMTF00123  AMTF00190  AMTE00158  AMTG00136  AMTD00108  AMTD00105  AMTF00121  AMTD00122  AMTE00220  AMTE00159  AMTG00138  AMTE00157  AMTE00156  AMTF00122  AMTF00124  AMTG00137  AMTD00110  AMTF00192  AMTD00109    d) 1006-9028-000  Lot/Serial Numbers:    AMTF00170    e) 1006-9310-000-305077  Lot/Serial Numbers:    AMTT00236    f) 1006-9310-000-015243  Lot/Serial Numbers:    AMTL00106    g) 1006-9310-000-017602  Lot/Serial Numbers:    AMTR00110  AMTL00285  AMTP00299  AMTL00194    h) 1006-9310-000-103785  Lot/Serial Numbers:    AMTR00120    I) 1006-9310-000-025109  Lot/Serial Numbers:    AMTM00216  AMTM00196    j) 1006-9310-000-009650  Lot/Serial Numbers:    AMTK00152    k) 1006-9310-000-015224  Lot/Serial Numbers:    AMTL00102    l) 1006-9310-000-031881  Lot/Serial Numbers:    AMTN00100  AMTN00101    m) 1006-9310-000-031854  Lot/Serial Numbers:    AMTN00211      n) 1006-9310-000-026571  Lot/Serial Numbers:",
      "more_code_info": ""
    }
  ]
}