  "results": [
    {
      "status": "Ongoing",
      "city": "Bolton",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90849",
      "recalling_firm": "Medicina Uk Ltd",
      "address_1": "Unit 2",
      "address_2": "Rivington View Business Park",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0112-2023",
      "product_description": "Medicina Reusable Oral syringe 1ml  Code: OTH01",
      "product_quantity": "111 boxes (11,100 syringes)",
      "reason_for_recall": "Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification",
      "recall_initiation_date": "20220704",
      "center_classification_date": "20221020",
      "report_date": "20221026",
      "code_info": "UDI-DI:  (01)05056115402869 (01)05056115402869(17)260700(10)HM21053   Batch number: HM18041, HM21053",
      "more_code_info": ""
    }
  ]
}