  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88692",
      "recalling_firm": "Uoc Usa Inc",
      "address_1": "15251 Alton Parkway Ste 100",
      "address_2": "N/A",
      "postal_code": "92618",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide:  CA, CO, and NV    No O.U.S.",
      "recall_number": "Z-0110-2022",
      "product_description": "United Orthopedics USTAR II Knee System    USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025",
      "product_quantity": "3 devices",
      "reason_for_recall": "Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.",
      "recall_initiation_date": "20210827",
      "center_classification_date": "20211013",
      "termination_date": "20230120",
      "report_date": "20211020",
      "code_info": "Catalog Number: 2915-1025  UDI Code: 04719872148197  Lot Code: 19E272H"
    }
  ]
}