  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93071",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "N/A",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0109-2024",
      "product_description": "THUNDERBEAT, 5MM, 45CM, Front-actuated Grip Type S  Model: TB-0545FCS  Indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc) and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed",
      "product_quantity": "271 units",
      "reason_for_recall": "Thunderbeats blue seal button may remain in the engaged position after the button is released and not immediately return to a neutral position may result in prolonged surgery",
      "recall_initiation_date": "20230912",
      "center_classification_date": "20231012",
      "report_date": "20231018",
      "code_info": "UDI-DI: 4953170409684 Lot Numbers: KR310806  KR310809  KR313734  KR319658  KR326619  KR326628  KR333661   Lots Added January 11, 2024: KR310806 KR310809 KR333661 KR342666"
    }
  ]
}